14 

 

 

 

 

5 Connection and Setup 

5.0

 

Ground Connections 

 

First make sure to connect the unit to a proper ground through the ground screw located on the 

left side of the oil tester. 

 

5.1

 

Prepare Oil Test 

 

Make sure the oil cup is properly cleaned twice before each oil sample test to ensure accurate 

results.  Fill  the  oil  cup  with  the  specified  amount  according  to  the  standard  being  tested  to.  Carefully 
place the oil cup into the slot holders of the transformer inside the test chamber. 

Summary of Contents for OC-DI Series

Page 1: ...I Operating Instructions OC DI Series Digital Interface Oil Testers Version 1 0 No 145896...

Page 2: ...II Title Digital Interface Oil Testers Date 06 2018 Revision History V0 1 03 2018 MC Initial draft of the document...

Page 3: ...protection of the system and will reduce its resistance to interference and transients It may also cause the individual unit to be no longer compliant with the relevant EMC emission and susceptibility...

Page 4: ...tions are used Indicates hazards There is a risk of equipment damage or personal injury or death Carefully read and follow the instructions Be sure to follow any safety instructions given in addition...

Page 5: ...provides a solution which achieves the optimal combination of traditional know how and leading edge technology This operating manual is designed for completeness and easy location of the required info...

Page 6: ...mmendations 10 2 2 Summary 11 3 Theory 12 3 0 Introduction 12 3 1 Pre Loaded Test Standards 12 3 2 Application 12 4 Oil Tester Ports 13 4 0 Ports 13 5 Connection and Setup 14 5 0 Ground Connections 14...

Page 7: ...Storage 23 7 2 Packing and Transport 24 7 3 Recycling 24 8 Customer Support 25 9 Appendix 26 9 0 Declaration of Conformity 26 9 1 Hipotronics Inc 26 9 2 Warranty 27 9 3 Returned Material RMA Error Bo...

Page 8: ...ipment is incomplete or damaged then this must be reported immediately to the transport company and the Customer Support Department of HIPOTRONICS or the local agent Repair or replacement of the instr...

Page 9: ...ure 4 158 F 20 70 C Humidity 5 95 r h non condensing Dimensions W x D x H OC60 DI 16 x 13 x 15 41cm x 33cm x 38cm Weight OC60 DI 60lbs 25kg 1 3 2 Power Specifications Power Input 90 264VAC 50 or 60Hz...

Page 10: ...d by their company and government Local and state safety procedures should also be consulted Company regional or national regulations must be fulfilled beyond HIPOTRONICS recommendations If the instru...

Page 11: ...ment involve personnel who are not familiar or perhaps too familiar with high voltage equipment Staying alert and ever watchful requires constant training and awareness of the inherent hazards The gre...

Page 12: ...n tests using Basic mode An internal kilovoltmeter automatically records the breakdown voltage for each test sample Each test can be saved into the unit s internal memory or to a USB drive 3 1 Pre Ins...

Page 13: ...or recharge unit When charging make sure that the input switch is in the ON position Ground Terminal Connect chassis to proper ground Power Button Turn on off unit interface When the unit is on the po...

Page 14: ...on the left side of the oil tester 5 1 Prepare Oil Test Make sure the oil cup is properly cleaned twice before each oil sample test to ensure accurate results Fill the oil cup with the specified amoun...

Page 15: ...center the temperature and humidity will be shown and in the right corner will be the battery level status and the current date time There are four windows that you can access from the Main Menu windo...

Page 16: ...voltage in one test sequence The ramp rate will set how fast the voltage increases during the test The wait time is the period between each individual breakdown test and the dwell time is the period t...

Page 17: ...ment in the top left will show the real time voltage across the electrodes The dwell and wait time counters will always be displayed and will count down when they are active The total elapsed test tim...

Page 18: ...the parameters for that step To remove the last step in the sequence click the x button in the bottom left of the table The sequence table will only ever show three steps at a time to view the other...

Page 19: ...the first step s target voltage at the selected ramp rate and then dwell at that voltage for the selected dwell time Once that step s dwell expires the voltage will start ramping up to the next step...

Page 20: ...test parameters will be displayed along with the calculated average breakdown voltage standard deviation of the breakdowns the breakdown range and a graph of the breakdowns The user can also export th...

Page 21: ...m s settings The user can either choose to default load the system presets when the unit is turned on or use the parameters from the most recently performed test The adjustable system parameters inclu...

Page 22: ...ains Unit is still able to be operated while charging without any adverse effects Lid Open While the lid cover is not over the high voltage chamber the test will not continue and high voltage is preve...

Page 23: ...ser alcohol or spirits Caustic cleansers and solvents Trio Chlorothene etc should definitely be avoided 8 0 2 Instrument Calibration When delivered new from the factory the instrument is calibrated in...

Page 24: ...better should be used Whenever possible protect the instrument from mechanical damage during transport with padding Mark the container with the pictogram symbols Fragile and Protect from moisture 8 3...

Page 25: ...reached at the following address HIPOTRONICS Customer Service 1650 Route 22 Brewster NY 10509 Tel 1 845 279 3644 E mail service hipotronics com We prefer contact via email Then the case is documented...

Page 26: ...he requirements of the listed standards or other normative documents So the product complies with the requirements of the EMC directive 2004 108 EC and the low voltage directive 2006 95 EC Product OC6...

Page 27: ...icing its position as to liability purchaser in processing or settling any such claim without prejudicing its position as to liability Warranty Claim Procedure Compliance with the following Warranty C...

Reviews: