![background image](http://html1.mh-extra.com/html/hipotronics/oc-di-series/oc-di-series_operating-instructions-manual_2137719022.webp)
22
7 Troubleshooting
7.0
Diagnostics
7.1
Warning and Error Messages
“Low Battery Please
Recharge the battery
”
The internal battery is low and needs to be recharged via AC
Mains. Unit is still able to be operated while charging without
any adverse effects.
“Lid Open”
While the lid cover is not over the high voltage chamber, the
test will not continue and high voltage is prevented to be
present.
Screen Flickering upon
start-up
The unit’s internal battery is too low and cannot complete the
start-up sequence. Please recharge the unit via AC Mains.