8. To output data using a Mitutoyo Input Tool, set 

the No. 3 connector of the DIP switch (page 6) 

to the on position (so the "P.ON" LED lights), 

set the

MODE

switch to "PEAK" and press 

the

RESET

switch (which functions the same 

as the Enter key on a computer) . The Input Tool 

will convert the measurement data for direct 

input to a spreadsheet application. In order to 

output data, please conduct measurement within 

the instrument's peak measurement range, as in-

dicated in the specifications. 

  

 Minus signs denoting counterclockwise torque 

are transferred with the measurement data. 

9. Fit the bit into a fastened screw to be tested. 

Clockwise rotation measures tightening torque 

and counterclockwise rotation measures loosen-

ing torque. All torque measurements are dis-

played digitally and the data is stored.

10. Be sure to turn off the power when you are finished 

measuring.

  Operating instructions

First check the state of the battery charge. When 

the power is on and the battery has an insufficient 

charge, the display reads "LOBAT." When this hap-

pens, use the battery charger to recharge the battery 

for at least 3 hours and no more than 6 hours. 

1. Connect the sensor and display units using the 

sensor unit cord. The plugs on either end of this 

cord have the same pin arrangement, so it doesn’t 

matter which unit gets which plug, but one is L-

shaped and the other is straight, so choose the 

combination most convenient for operation. 

When making the connection, look at the pins in 

the plug and take care to fit them smoothly into 

the connector to avoid bending them.

2.  Attach the bit provided with the torque meter to 

the sensor unit.

Fit the joint shaft collar over 

the driver and insert the bit 

into the collar. 

3.  Select  the  unit  of  measurement  desired

UNIT

 switch. (N

m

N

cm) (N

m

lbf

in) 

(N

m

kgf

cm)

4.  Conduct zero adjustment. Set

MODE

 switch 

to "TRACK," then press both the 

SET

and 

【▲】

switches at the same time for less than 2 

seconds. This adjusts the meter to zero. (Note: 

If you press the switches for than 2 seconds, the 

meter will go into another mode. If this happens, 

press

RESET

, then repeat the zero adjust-

ment process.)

5. Set

MODE

switch  to "PEAK" or "F.PEAK."

 

 •

 In peak mode the peak torque value reached dur-

ing measurement is displayed. Thus, the display 

does not change. Model HDP-50 measures up to 

10 N

m of torque, which corresponds to a dis-

play value of 15. Please conduct measurements 

within this range. 

 

 •

 In "F.PEAK" mode the first peak value is dis-

played and does not change even if higher torque 

is applied.

6.  Press the

RESET

switch to cancel the display.  

7. Minus signs appear in front of measurements of 

counterclockwise torque. 

-  5  -

Summary of Contents for HDP-100

Page 1: ...17A HIOS Digital Torque Meters HDP 50 HDP 5 HDP 100 Option Operation manual May 2017 Operation manual No ET C007 Utilizing 100 post consumer recycled paper pulp ...

Page 2: ...ting screws by required torque values Classifying jars by screw cap tightness Destructive testing of small parts Quality control of plastic screw threads and plastic parts Features Up to 100 data measurements can be stored in memory at a time Checks maximum minimum and average torque measurements Torque value display can be automatically reset to zero Torque value display are automatically recali ...

Page 3: ...120 will result in internal damage to the meter The battery charger is for NiMH batteries only Nev er use any other kind of battery even from HIOS Do not disassemble the instrument or loosen the screws on its body as this could cause inaccu racy in measurement Protect the display and sensor units from shocks or impacts Always detach the bit from the sensor unit after use Use of the battery charger...

Page 4: ...or the coun try not described in the list page 4 The conversion plugs attached with TCH MULTI CE area will be no longer available as soon as the current stock has run out We would like to ask those who use BF and S plug to get them locally 220 Volt Adaptor Plug Market Areas China Japan Korea Taiwan Hong Kong Singapore Europe U S A Canada U K Australia HDP 50 HDP 5 HIOS shank 5 Bit 0 60 mm 1 60 mm ...

Page 5: ...mory function toggle switch The LED lights when the memory function is on M_SET Torque set switch B Sets the torque level at which the warning tone sounds Max load warning tone The LED lights when this function is on on off switch BZ Memory clear switch M_CR Deletes data in memory Average measurement When average measurement mode is on the LED lights and the torque measure mode switch AVE ment ave...

Page 6: ...lugs on either end of this cord have the same pin arrangement so it doesn t matter which unit gets which plug but one is L shaped and the other is straight so choose the combination most convenient for operation When making the connection look at the pins in the plug and take care to fit them smoothly into the connector to avoid bending them 2 Attach the bit provided with the torque meter to the s...

Page 7: ...ched This tone also sounds when the torque set ON ON Reset Setting Warning tone Track Peak Switch functions for setting torque Switches used SET switch 1 Press for 2 seconds to enter torque setting mode Press for 2 seconds once again to exit torque set ting mode 2 When the meter is in torque setting mode the 4 digit display can be adjusted one digit at a time Use the SET switch to select the digit...

Page 8: ... position 0001 0002 0003 0004 0001 0002 0003 0004 Fig 4 Full deletion of measurement data Partial deletion of measurement results Example of full deletion of measurement data Memory function on LED on measurement data in memory Ready for measurement M_CR switch M_CR switch Full data deletion mode display Press for 2 seconds Display flickers Exit full deletion mode RESET pressed or no action for 30...

Page 9: ...ues are displayed flickering Fig 1 Min Ave 1st measurement deleted 11th measurement recorded 0001 0002 0003 0010 0001 0002 0003 0009 0010 0001 0002 0003 0009 0010 Notes 1 Be sure to set MODE switch to PEAK or F PEAK before entering average measurement mode 2 The UNIT switch does not function in average measurement mode Battery charger To charge batteries for the HIOS torque meter use only the batt...

Page 10: ...ice must be paid by the customer Please direct questions about customer services to HIOS Corporation or your HIOS dealer Attention The product that you have purchased contains a rechargeable battery The battery is recyclable At the end of it s useful life under various state and local laws it may be illegal to dispose of this bat tery into the municipal waste stream Check with your local solid was...

Page 11: ...duct zero adjustment The MODE switch is set to TRACK The battery charger plug may not be en gaged all the way into its socket The connection may be to the wrong termi nal The battery may have reached the end of its operational life The battery charge may be insufficient This may be due to noise occurring from the power source etc Troubleshooting Before you conclude the torque meter is broken go th...

Page 12: ...ignments Mitutoyo MQ65 5P GND Ground DATA Outputs data in the for mat indicated below CK Clock RD Demands data reception 2 Data output format Following is an explanation of the meanings of the 13 digit data output from the torque meter Sign Entry number Data type Decimal point position Measured value REQ Demands data output from exter nal source Open drain 0 3 7 V 400 µA max Pull up to VDD 1 55 V ...

Page 13: ...oduct and product box At the product you can find it at the bottom and it is marked on the product box If you cannot find the mark please ask your distributor In case of emergency please cut the mark below and stick at the bottom of product and on the product box China RoHS mark Note minimum size of China RoHS mark is 5mm at outer diameter 12 ...

Page 14: ......

Page 15: ......

Page 16: ...1 16 5 Akiyama Matsudo City Chiba Pref Japan TEL 81 Japan 47 392 2001 FAX 81 Japan 47 392 7773 ...

Reviews: