The display says "
LOBAT
"
There is no display
The displayed value does
not return to zero
The displayed values do
not stop changing
Unable to recharge the
battery
The "
LOBAT
" display reap-
pears after battery recharg-
ing is complete
The display shows a
numerical value when it
shouldn't
Problem
Causes
Resolving the problem
•
Recharge the battery as explained on page 8.
If this
does not resolve the problem, contact HIOS or your
HIOS dealer.
•
Turn the power switch off, conduct a full battery
recharge (6 hours, no more), then turn the power
switch on and see if there is a display. If there is not,
contact HIOS or your HIOS dealer.
•
Conduct zero adjustment.
•
Set to "
PEAK
."
•
Check that the plug is connected properly.
•
Connect to the battery charger socket.
•
If the above does not remedy the problem, contact
HIOS or your HIOS dealer.
•
Send to HIOS for repair.
•
Recharge the battery for no more than 6 hours.
•
Press
【
RESET
】
switch to clear the display.
•
Insufficient battery charge
•
This may be due to long-term use of the
meter or natural battery discharge over
time.
•
Unable to conduct zero adjustment.
•
The
【
MODE
】
switch is set to "
TRACK
"
•
The battery charger plug may not be en-
gaged all the way into its socket.
•
The connection may be to the wrong termi-
nal.
•
The battery may have reached the end of
its operational life.
•
The battery charge may be insufficient.
•
This may be due to noise occurring from
the power source, etc.
■
Troubleshooting
Before you conclude the torque meter is broken, go through the following troubleshooting guide. If, by fol-
lowing the instructions, you are unable to resolve the difficulty, please contact HIOS Corporation or your
HIOS dealer.
Peak range
N•m
0.015 - 0.500
0.15 - 5.00
0.15 - 10.00
N•cm
1.5 - 50.0
15 - 500
15 - 1000
lbf•in
0.15 - 4.40
1.5 - 44.0
1.5 - 90.0
(kgf•cm)
0.15 - 5.00
1.5 - 50.0
1.5 - 100.0
Accuracy*
within ±0.5% (at full scale)
Power
five 1.2 V NiMH cells 1,000 mA
Recharge time
up to 6 hours
Running time on full charge
8 hours
(continuous operation)
Battery life
About 300 chargings
Dimensions
See figure
Weight
Display
780 g
Sensor unit
68 g
250 g
Battery charger
Runs on 100 - 240 V AC power (50 or 60 Hz)
Output 7.3 - 8 V DC (0.5 A)
Model
HDP-5 HDP-50 HDP-100
■
Specifications
• Never exceed the maximum
torque tolerance level for the in-
strument.
• Because battery life depends on
operating conditions, this cannot
be guaranteed.
- 10 -