51
(15) Others
Set and Query Power LED Lighting
Syntax
Command
:SYSTem:POWer:LED
<1/0/ON/OFF >
Query
:SYSTem:POWer:LED?
Response
<1/0(NR1)>
Description
Command
Sets the power LED lighting.
ON: Lights up or flashes in accordance with the instrument state.
OFF: Always off.
Query
Queries the currently set power LED lighting setting.
Example
:SYST:POW:LED ON
:SYST:POW:LED?
1
Note
• This setting is not stored. It starts in a state of ON at power-on.
• When OFF is set, an error is not notified using the power LED. Use
*ESR?
or
:SYSTem:ERRor?
to acquire the error information periodically.
Execute System Reset
Syntax
Command
:SYSTem:PRESet
Description
Resets the operation of
*RST
and the communication setting.
Example
SYST:PRES
*OPC?
1
Note
Be sure to check that the operation is complete using
*OPC?
. When the communication
disconnection process is executed without checking the operation completion, the
initialization process may not be executed.
However, when the communication mode setting is the user setting mode, the
communication is disconnected as the reset process is complete. At this time, the
operation completion cannot be checked using
*OPC?
.
When turning off the power after this command has been executed in the user setting
mode, turn off the power after 1 sec. or longer has elapsed.
When you turn off the power immediately after this command has been executed,
the initialization process may not be executed.