![Hioki MR8880-20 Instruction Manual Download Page 231](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332231.webp)
12.3 External Signal Output (Trigger Output)
221
Chapter 12 Ext
ernal Cont
rol
12
You can output a signal when a trigger event occurs. This allows synchronous oper-
ation of multiple Memory HiCorders by parallel triggering (p. 226).
For the external trigger setting procedure, see "Applying Trigger with External Input
Signal (External Trigger)" (p. 127)
12.3 External Signal Output (Trigger Output)
Pulse wave is output.
Trigger event occurs
TRIG.OUT
Trigger Output Signals
Output signal
Open drain output (with voltage output), Active Low
Output voltage range
HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Pulse width
Level: Sampling period x (number of data since trigger –1) or
longer (2 µs or longer) (for trigger synchronization operation)
Pulse: 2 ms ±10% (for external synchronization)
Maximum switch rating
5 to 30 VDC, 200 mA
10 k
Ω
5 V
TRIG.OUT
GND
HIGH
4.0 to 5.0 V
LOW
0 to 0.5 V
2
μ
s or greater
• When triggering is not otherwise used, a trigger signal is output during
measurement.
A trigger signal is also output when the Auto Range function is used,
so be careful when using both Auto Range and the trigger output sig-
nal for triggering other devices.
• When using an external trigger to synchronize multiple instruments,
make sure you set the external trigger output to
[Level]
on the Sys-
tem screen of the external trigger input.
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......