174
Detailed Settings for Modules
• The calibration time is approximately 150 ms. During this period measurement is not performed.
• If synchronization between channels is performed, a signal to stop integration is sent to each module
when measurement is started and the wait process is performed until the first integration finishes.
The wait time required for this process is (10 ms + integration time*).
* The integration time varies depending on the NPLC setting.
If synchronization is not performed, the above wait time will also be required for measurement
performed immediately after the settings of the Model MR8990 Digital Voltmeter Unit are changed.
However, there will be no wait time when measurement is performed with the same settings.
• When this is set to
[Off]
(Default setting), perform calibration manually.
Refer to “2.8 Perform Calibration (When Model MR8990 is installed)” (p. 48).
8.10.9 Setting Model U8974 High Voltage Unit
and Make a Channel Selection” (p. 164)
Mode
Switches between voltage measurement and RMS measurement.
DC
Voltage measurement (Default setting)
RMS
RMS measurement
Response
The response time for RMS measurement can be set to three speeds: Fast, Normal
and Slow.
The response time can be set to
[Slow]
to stabilize the measured value when the
frequency is low or when severe fluctuations are present.
Fast
Response time is set to 150 ms.
Normal
Response time is set to 500 ms. (Default setting)
Slow
Response time is set to 2.5 s.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......