Page 1: ...dition 1 IM9202A961 00 20 04H Be sure to read this manual before using the instrument p 3 When using the instrument for the first time Troubleshooting Part Names and Functions p 6 Maintenance and Serv...

Page 2: ...HIOKI IM9202A961 00...

Page 3: ...t how to use the product In explaining how to use the product it assumes electrical knowledge equivalent of the knowledge possessed by a graduate of an electrical program at a technical high school Co...

Page 4: ...Instruction manual The H side lead component clip and the L side lead component electrode are installed on the device Options sold separately The options listed below are available for the device To...

Page 5: ...h of or serious injury to the operator CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the device or malfunction IMPORTANT...

Page 6: ...ce Handling the Device CAUTION The DUT could pop off the device and into the air while being connected To prevent injury use gradual movements when handling clips and electrodes To avoid damage to the...

Page 7: ...Lead wire diameter 0 5 mm to 2 mm Radial D E D Lead interval 2 mm to 26 mm E Lead length 1 2 mm or longer Lead wire diameter 0 5 mm to 2 mm SMD 2 H F G F Component length 1 6 mm to 23 mm G Component...

Page 8: ...on The exclusive stand for fixing the test head of the impedance analyzer and the device 7 Loupe The magnifying glass provided with the IM9200 Test Fixture Stand option It is used by installing to the...

Page 9: ...d components 10 9 2 The electrode for measurement of SMDs 13 14 13 L side SMD electrode Used when measuring SMDs 14 Lever This lever causes the L side SMD electrode to slide back and forth Short plate...

Page 10: ...struction manual of the impedance analyzer Test head 2 Attach the test head and adapter Attach the test head and adapter to the stand to the stand Attach the test head and the IM9906 Adapter to the IM...

Page 11: ...values fall within the allowable range See Residual parameters reference value p 24 Check items Remedy Are the electrode or the short plate worn If any of the following applies send the device for re...

Page 12: ...9906 Adapter 1 3 5 mm connector torque wrench 1 not included with the device Test head IM9200 1 1 2 2 Metal fittings 1 Pull the metal fittings to the IM9200 Test Fixture Stand side 2 Place the test he...

Page 13: ...f the internal conductor rotates it may cause damage to the electrode When the calibration cannot be executed Confirm that the calibration kit and the IM9906 Adapter are connected properly Check that...

Page 14: ...each other and the calibration kit and adapter may get damaged 6 Execute open calibration by using the impedance analyzer 2 2 1 1 0 0 3 3 4 4 Nut Short calibration kit 9 9 7 Remove the open calibratio...

Page 15: ...in figure connect by turning only the nut of load calibration kit while pressing the top part of the load calibration kit If the top part of the load calibration kit is rotated the internal conductors...

Page 16: ...1 Lower the screw thread by turning the nut of the IM9906 Adapter 2 2 2 2 3 3 2 Loosen the stand fixing knob 3 Place the device on the IM9200 Test Fixture Stand As shown in figure slide the notch part...

Page 17: ...he error caused by the phase shift occurring between the calibration reference surface and the measurement DUT s connection surface is compensated Set the electrical length correction to 0 mm in the i...

Page 18: ...he difference in impedance between the DUT and the short device The short device s impedance will affect measured values Using a fixed reference without regard to the size of the DUT The value after s...

Page 19: ...easurement Knob L side lead component electrode Base plate 2 If electrodes and clips for lead component measurement have been installed Remove the H side lead component clip and L side lead component...

Page 20: ...r When the correction cannot be executed Confirm that the IM9906 Adapter is connected properly Check that the DC measurement limit value of the impedance analyzer is appropriate 1 1 2 2 3 3 4 4 5 5 0...

Page 21: ...screwdriver install the H side lead component clip 2 Gently tighten the knob that holds the L side lead component electrode in place Edge 3 3 H side electrode Marker 3 Slide the L side lead component...

Page 22: ...hoose the best L side SMD electrode lock position from 1 through 4 as shown in the figure on the left based on the SMD s dimensions width Lock position for the L side SMD electrode SMD dimensions widt...

Page 23: ...the plate Catch the other protruding part on the leading edge of the L side SMD electrode View from the side with the jig installed SMD open correction jig If you re unable to secure electrode positi...

Page 24: ...er on the L side electrode to sandwich the component between the H side electrode and the L side SMD electrode Verify that the edge of the L side SMD electrode is in contact with the base plate Proper...

Page 25: ...ation Connecting part to the DUT 2 terminal structure Lead components The lead component s leads are pinched between the device s electrodes and clips SMDs The SMD is pinched between the device s elec...

Page 26: ...y connected to the H side electrode The H side clip is present 1 2 nH 1 1 nH 30 MHz 1 2 nH 0 8 nH 100 MHz 1 2 nH 0 7 nH 300 MHz 600 MHz Residual parameters Short inductance L side SMD electrode Condit...

Page 27: ...d for the duration of the recommended replacement cycle Parts Recommended replacement cycle Conditions Base plate 10000 times Number of times the L side lead component electrode and L side SMD electro...

Page 28: ...26 Maintenance and Service HIOKI IM9202A961 00...