9
Making Measurements
Inspecting the Device Before Use
To ensure that the device is properly operating, conduct an inspection and check device operation
before using device to ensure that no damage has occurred during storage or transport. If there is
any damage to the device, contact your authorized Hioki distributor or reseller for repair.
It is recommend to inspect the device prior to use as described below.
Measure residual parameters with an impedance analyzer and verify that the measured values fall
within the allowable range.
See "Residual parameters (reference value)" (p. 24)
Check items
Remedy
Are the electrode or the short
plate worn?
If any of the following applies, send the device for repair:
•Residual parameters’ measured values diverge from the allowable
values.
•An electrode or the short plate exhibits wear.
Does either of the clips exhibit any
damage or unusual appearance?
If any of the following applies, send the device for repair:
•Either of the clips is cracked or worn.
•Either of the clips will not move.
•The clips are not able to accommodate a component.
Is the device damaged?
Send the device for repair if it exhibits any deformation in shape or
other damage.