Recording Capacity

• Maximum recording time is inversely proportional to number of recording analog channels.
• Because the actual capacity of a CF card is less than that indicated, , expect actual maximum times to be about 90% of those in the table.
• "

" exceeds 1 year.

• Proper operation is not guaranteed for extended recording periods (one year or longer). This type of operation impacts the product’s warranty period and service life.

Investigate why your office’s power sup

-

ply occasionally exhibits instability.

Check for fluctuations in low-voltage 

signals such as instrumentation or sen-
sor output.

Measure the phase voltages for all 
three phases of a three-phase motor 
simultaneously.

Four channels of isolated 
CAT III 600 V input

Thanks to its 14-bit, high-resolution A/D converter and the combina-

tion of a high-sensitivity 10 mV/div range and a 5 Hz filter (for noise 

rejection), the MR8880 can deliver stable measurement of sensor out-
put. 

The MR8880 is capable of mixed recording of RMS values, DC 
voltage, and logic signals, allowing it to simultaneously record 
data describing the interrelationships between equipment power 
supplies and UPS output and control signals. 

The 

MR8880

 provides a turnkey solution for both high-speed measurement at 1 MS/s and long-term measure-

ment. Its ability to measure everything from high- to low-voltage signals allows it to play an important role in a 
variety of measurement scenarios.

Recording Time

 (Internal memory)

All channels (4 8 logic channels)

Time Axis Range

Sampling Speed

Recording Interval Max. Recording Time

100

μ

s/DIV

1 MS/s

1 μs

1 s

200

μ

s/DIV

500 kS/s

2 μs

2 s

500

μ

s/DIV

200 kS/s

5 μs

5 s

1ms/DIV

100 kS/s

10 μs

10 s

2

ms/DIV

50 kS/s

20 μs

20 s

5ms/DIV

20 kS/s

50 μs

50 s

10ms/DIV

10 kS/s

100 μs

1m 40 s

20ms/DIV

5 kS/s

200 μs

3m 20 s

50ms/DIV

2 kS/s

500 μs

8m 20 s

100ms/DIV

1 kS/s

1 ms

16m 40 s

The maximum recording length is fixed regardless of the number of channels in use.

Measure RMS value fluctuations for a power 

line over an extended period of time

Long-term measurement and recording us-
ing the real-time function

Recording 

Interval

All channels (4 8 logic channels), recording waveform (binary) data only

Internal memory

(8MB)

512MB
 (9728)

1GB

(9729)

2GB

(9830)

100

μ

s

1m 40s

1h 25m 20s

2h 46m 40s

5h 33m 20s

200

μ

s

3m 20s

2h 50m 40s

5h 33m 20s

11h 6m 40s

500

μ

s

8m 20s

7h 6m 39s

13h 53m 19s

1d 3h 46m 39s

1ms

16m 40s

14h 13m 19s

1d 3h 46m 39s

2d 7h 33m 19s

2

ms

33m 20s

1d 4h 26m 38s

2d 7h 33m 18s

4d 15h 6m 38s

5ms

1h 23m 20s

2d 23h 6m 34s 5d 18h 53m 14s 11d 13h 46m 34s

10ms

2h 46m 40s

5d 22h 13m 8s 11d 13h 46m 28s

23d 3h 33m 8s

20ms

5h 33m 20s 11d 20h 26m 15s 23d 3h 32m 55s

46d 7h 6m 15s

50ms

13h 53m 20s

29d 15h 5m 39s 57d 20h 52m 19s 115d 17h 45m 39s

100ms

1d 3h 46m 40s

59d 6h 11m 17s 115d 17h 44m 37s 231d 11h 31m 17s

200ms

2d 7h 33m 20s 118d 12h 22m 34s 231d 11h 29m 14s

-

-

500ms

5d 18h 53m 20s 296d 6h 56m 26s

-

-

1s

11d 13h 46m 40s

-

-

2s

23d 3h 33m 20s

1 min

694d 10h 40m

-

-

-

-

-

-

Recording interval of 100 μs to 1 min

Waveform data is saved directly in a binary format to a CF card or 

USB memory.

Measure the instantaneous wave-
form at startup or a suddenly gen-
erated abnormal waveform.

High-speed measurement using the high-speed function

Fastest sampling period of 1 μs (measuring all channels simultaneously)

Measurement data is recorded in the instrument’s internal memory (1 MW).

1

2

3

4

5

Record RMS values and 

instantaneous waveforms 

simultaneously.

Example record of an abnormal 
waveform

Applications

Note: Use only Hioki CF cards that are guaranteed to operate with the HiCorder 

for continuous long-term recording.

A  waveform  recorded  using  a  waveform 
judgment  trigger. The  judgment area can 
be displayed simultaneously.

The MR8880 can measure the voltages 
at different contacts without the need for 
a differential probe. 

Summary of Contents for HiCORDER MR8880

Page 1: ...ions to select the appropriate settings The re corder can be easily configured to measure voltage drops and power outages Maximum 600V AC DC input no need for a differential probe 4 completely isolate...

Page 2: ...22 F Even when operating on battery power the MR8880 can take measurements from 0 C 32 F to 40 C 104 F n Rugged damage resistant design features standard side protectors that guard the instrument s ca...

Page 3: ...the channel you wish to use Select the power line voltage Select the frequency Select the threshold Select the high speed or real time function The auto range settings can be enabled when using the hi...

Page 4: ...kS s 500 s 8m 20 s 100ms DIV 1 kS s 1 ms 16m 40 s The maximum recording length is fixed regardless of the number of channels in use Measure RMS value fluctuations for a power line over an extended pe...

Page 5: ...aking measurements in an outdoor setting The dis play is even engineered for easy viewing in the presence of reflections a What if there s no power available in the vehicle being tested A high capacit...

Page 6: ...hannels isolated input and outputs logic has common GND Maximum sampling rate 1Msamples s 1 s cycle all channels simultaneously Memory capacity 14bit 1 M words ch 1 word 2 bytes not expandible Externa...

Page 7: ...open and 8 k or lower short Response speed 500ns or lower Max allowable input 0 to 50V DC the maximum voltage that can be applied across input pins without damage LOGIC PROBE MR9321 01 Function Detec...

Page 8: ...06 ft roll type 10 rolls set Other options CARRYING CASE C1003 Includes compartment for options soft case type CONNECTION CORD L9217 Cord has insulated BNC connectors at both ends for signal output 1...

Reviews: