34
Inspecting Before Use
DANGER
•
Connect the instrument to the secondary side of
a distribution panel. If a short-circuit occurs on
the secondary side of the distribution panel, the
panel will interrupt the short-circuit current. Do not
connect the instrument to the primary side of the
distribution panel because an unrestricted current
flow can damage the instrument and facilities if a
short-circuit occurs.
•
Do not cause a short-circuit between another wire
and the wire to be measured with the test leads.
Arcs or such grave accidents are likely to occur.
•
To prevent a short-circuit or electric shock, do not
touch the metal part of the connecting test leads tip.
WARNING
•
To prevent an electric shock, confirm that any of the
white or red portion (insulation layer) of the L9787
Test Lead and the L9844 Measurement Cable is not
exposed. If the cable is exposed, do not use the
cable.