5. Detail of Commands
5.1. Description of Commands
The DT4251,DT4252,DT4253,DT4254,DT4255,DT4256 only accepts the UPPERCASE
command. (Except Unit etc.)
Table.4
Command
Explanation
QPID
Query the Meter model.
Syntax QPID
Response “DT4261”
*IDN
Query the Meter identification.
Syntax *IDN?
Response <data×3>
<data> maker name, model number, serial, version
Example; “HIOKI,DT4261,210601234,Ver 1.00”
:SYST:RST
Put the meter to reset.
Syntax :SYST:RST
Response “OK”
:SYST:LLO
Put the meter into the local lockout state.
Syntax :SYST:LLO
Response “OK”
:SYST:GTL
Put the meter into the local state.
Syntax :SYST:GTL
Response “OK”
:CONF?
:CONF2?
Query the function of the display.
Syntax :CONF? [:CONF2?]
Response <data×2>
<data> function,range
Example; “DCV, 600m”
5.2 Combination response string function and range
:CONF
Configure the range.
Syntax :CONF <data×2>
<data> function,range
Example; “:CONF RES, 60k”
5.2 Combination response string function and range
Response “OK” or “CMD ERR” or “EXE ERR”