91
Errors
4
IMPORTANT
•
The instrument may be unable to properly detect overload
states immediately after the current range has changed.
•
The currents for which an overload state can be detected
are DC and sine waves with frequencies of 45 Hz to 66 Hz.
The device is unable to detect the currents listed below as an
overload state.
- Currents that exceed the defined level on a momentary
basis
- High-frequency currents that exceed the defined level
•
Although external magnetic fields may cause the
OVERLOAD
LED to blink while the upper jaw is retracted, this does not
indicate an issue with the device.
HIOKI CT6710A961-01