54
Measuring Currents
6
Convert a voltage sensitivity of the waveform
measuring instrument into a current sensitivity.
Using the following formula can convert a voltage sensitivity
(unit: V/div) specified on the waveform measuring instrument
into a current sensitivity (unit: A/div).
S
I
=
S
V
/
R
O
S
I
:
Current sensitivity
(A/div)
S
V
:
Voltage sensitivity
(V/div)
R
O
: Output rate
(V/A)
Example
When your waveform measuring instrument with its voltage
sensitivity set at 10 mV/div measures a current that has an
RMS value of 1 mA
1
mA/div
10
mA/div
0.5 A range (10 V/A)
5 A range (1 V/A)
S
I
= (10 ×10
-3
) / 10
S
I
= (10 × 10
-3
) / 1
= 1 × 10
-3
(A/div)
= 10 ×10
-3
(A/div)