Repairs, Inspections, and Cleaning
4.1 Troubleshooting
Symptom
Verification and/or Solution
•
The instrument is indicating an
abnormal measured value.
•
Is the measured current value too small for the instrument’s
measurement range?
•
Wrap the wire around the jaw one or more times. Each additional
wrap of the wire will increase the measured value, so that wrapping
it once yields a measured value that is twice the actual value and
wrapping it twice yields a measured value that is three times the
actual value.
•
Is the tip of the jaw open?
•
Is the jaw damaged?
•
If the jaw is damaged or cracked, it will not be able to measure current
accurately. Send the instrument for repair.
•
Displayed values can frequently
fluctuate due to induction potential
even when no voltage is applied. This, however, is not a malfunction.
4 Repairs, Inspections, and Cleaning