54
Accuracy Table
PEAK MAX/PEAK MIN
Range
Accuracy guarantee range
(Resolution)
Accuracy guarantee
frequency range *
1
Measurement accuracy
6.000 V
0.00 V to ±12.00 V
(0.01 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg.±0.07 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg.±0.07 V
66 Hz < f ≤ 1 kHz
±1.5% rdg.±0.07 V
60.00 V
±3.0 V to ±120.0 V
(0.1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg.±0.7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg.±0.7 V
66 Hz < f ≤ 1 kHz
±1.5% rdg.±0.7 V
600.0 V
±30 V to ±1000 V *
2
(1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg.±7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg.±7 V
66 Hz < f ≤ 1 kHz
±1.5% rdg.±7 V
1000 V
±50 V to ±1000 V *
3
(1 V)
10 Hz ≤ f < 45 Hz
±1.5% rdg.±7 V
DC, 45 Hz ≤ f ≤ 66 Hz
±1.0% rdg.±7 V
66 Hz < f ≤ 1 kHz
±1.5% rdg.±7 V
*1: Frequency range of 10 Hz ≤ f < 20 Hz is designed value.
*2: Values of up to ±1200 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
*3: Values of up to ±1700 V are displayed, but accuracy is not defined for display values in excess of 1000 V
(which are provided as reference values).
Summary of Contents for CM4141
Page 2: ......
Page 4: ......
Page 40: ...34 Bluetooth Communications only for model CM4142...
Page 64: ...58 Accuracy Table...
Page 73: ......
Page 74: ......
Page 75: ......