20
_____________________________________________________________________________________________
______________________________________________________________________________________________
Measurement
accuracy
±
0.1%rdg.
±
0.065%f.s.
(
23
℃±
5
℃
, 80%RH or less
)
(Adjust including energizing voltage accuracy and the cable
configuration.)
Suitable transducer
Strain gauge type transducer (Bridge resistance: 350
Ω
〜
1.5
k
Ω
)
Connectors in use (TAJIMI ELECTRONICS CO.,LTD,
PRC03-23A10-7F)
Bridge voltage
3 Vrms, 3 kHz
±
10%, Sine wave
Zero ajustment
±
10%f.s. or less
Measurement range
1 mV/V/1.5 mV/V/2 mV/V
Display (SC=1)
1.0000/1.5000/2.0000
Effective input
range
5% to 110% (Display range: 0.1%
〜
130%)
Maximum allowed
voltage
200% of range
Measurement
accuracy
±
0.1%rdg.
±
0.1%f.s. (23
℃±
5
℃
, 80%RH or less)
Input resistance
200 k
Ω
±
5% (Differential input)
Measurement range
±
1.0000/
±
5.0000/
±
10.000
V
Effective input
range
1% to 110% (Display range: 0.1%
〜
130%)
Maximum allowed
voltage
±
20 V
Strain Gauge Input (chA Special Connector)
DC Input (BNC connector of chA/chB)
Pulse Input (BNC connector of chB)
Frequency measurement: 1 Hz to 100 kHz. Measurement accuracy depends
upon the frequency measurement accuracy of the 3194.
Maximum allowed voltage:
±
20 V