5.4 Converting Input Values (Scaling Function)
127
5
Chapte
r 5
Input Chan
nel Setting
s
Using the Model 8939 or 8960 Strain Unit
For the rated capacity and rated output, consult the calibration record of the sen-
sor to be used. Set as follows:
Example 2.
Using the 20 G rated capacity and a sensor with 1000 µV/V rated output,
display measured data in units of [G]
Setting Items
Setting Choice
Scaling
On
Format
Decimal
Units
G
Capacity
20 [G]
Output
1000 [
V/V]
(displays as 1k)
By using the Scaling function, signals
from the sensor are acquired as physical
values.
A/B cursors and gauges are displayed
and printed as physical (G) values.
See
"Applying Gauges" (p. 198)
"List and Gauge Settings" (p. 330)