9.1 Basic Specifications
255
9
Specifications
Chapter 9
9.1 Basic Specifications
Product Specifications
Measurement items
C (capacitance), D (dissipation factor tan
)
Measurement Frequency
120 Hz, 1 kHz
Frequency accuracy: ±
0.01% or less
Measurement Signal Level
(1) Constant voltage mode: 100 mV (only for Model 3504-60), 500 mV, 1 V
Measurement range
CV1 V: to 70
F range (measurement frequency: 1 kHz)
CV1 V: to 0.7 mF range (measurement frequency: 120 Hz)
Cv100 mV, CV500 mV: to 170
F range (measurement frequency: 1
kHz)
Cv100 mV,CV500 mV: to 1.45 mF range (measurement frequency: 120
Hz)
Signal level accuracy of
±
10%
±
5 mV
(2) Open terminal voltage mode: 100 mV (only for Model 3504-60), 500 mV,
1 V
Measurement range: other than above
Output resistance: 5
±
1
Signal level accuracy:
±
10%
±
5 mV
Guaranteed Accuracy Range
C : 0.9400 pF to 20.0000 mF
D : 0.00001 to 0.10000 (0.10001 to 1.99999 are reference values)
Measurement Range
C : 009.400 pF to 20.0000 mF (120 Hz, 10 range)
00.9400 pF to 2.00000 mF (1 kHz, 10 range)
Auto, manual (up, down)
Equivalent Circuit Mode
Series and parallel equivalent circuit modes
Auto, manual
Measurement Time
Representative value: 2.0 ms
(measurement frequency: 1 kHz, measure-
ment speed: FAST)
* The measurement speed differs depending on the measurement fre-
quency and measurement speed.
Measurement Speed
FAST, NORMAL, SLOW
Trigger Function
Setting of internal trigger or external trigger is possible
Zero Compensation
Open circuit and short circuit compensation is possible
Load Compensation
Measuring the sample to be used as the standard sample and then com-
pensating the measurement values is possible
Offset circuit compensation
Set value cannot be subtracted from the measurement value
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......