3.3 DC Integration
41
1
2
3
4
5
6
7
8
9
10
B
asi
c M
easu
rem
e
nt
Proce
dure
3.3 DC Integration
Sensor used
: Model 9691 CLAMP ON AC/DC SENSOR
Measured current : 20 A or less
Manually stop measuring (no timer)
1.
Press the
button to select DC
(
).
Button presses are ignored if
TIME
is dis-
played: press
→
to clear it.
2.
Press
button once to select the
L range (20 A).
3.
Press
button to display the inte-
gral.
Unless output is required, the integrating
range can be left as AUTO.
4.
Press
SHIFT
→
(
0 ADJ
) to execute
automatic zero adjustment.
is displayed.
5.
Clamp around a conductor to be mea-
sured so that it is centered in the
clamp.
The figure shows the 9691 CLAMP ON AC/
DC SENSOR.
The current direction mark on Models 9692
and 9693 CLAMP ON AC/DC SENSORs is in
a different location.
6.
Press
SHIFT
→
(
START/STOP
) to
start integral measurement.
START
and
TIME
are displayed.
7.
Press
SHIFT
→
(
START/STOP
) to
stop integral measurement.
STOP
is displayed.
SOURCE
Electric
conductor
LOAD
The 9691 CLAMP ON AC/DC SENSOR
1
2
Position the clamp
with the current di-
rection indicator
pointing toward the
load side.