![Hioki 3280-1 OF Instruction Manual Download Page 27](http://html1.mh-extra.com/html/hioki/3280-1-of/3280-1-of_instruction-manual_2137187027.webp)
EN-19
Specifi cations
4
AC Current - Flexible loop (3280-20F)
Range
Accuracy
range
Accuracy
40 Hz
≤
f <
50 Hz
50 Hz
≤
f
≤
60 Hz
60 Hz < f
≤
1 kHz
420.0 A
40.0 A to
419.9 A
±3.5% rdg.
±5 dgt.*
1,
*
2
±3.0% rdg.
±5 dgt.*
1
±3.5% rdg.
±5 dgt.*
1,
*
2
4200 A
400 A to
4199 A
*1: Includes accuracy of CT6280 AC Flexible Current Sensor:
±1.0% rdg.
*2: Accuracy is not defi ned for currents of 1000 A or more or
currents of 5×10
5
A·Hz or more.
5
AC Voltage
Range
Accuracy
range
Accuracy
Input
impedance
45 Hz
≤
f
≤
66 Hz