EN-15
Specifi cations
Options
• CT6280 AC Flexible Current Sensor
(Attachment is included)
• 9209 Test Leads Holder
• L4933 Contact Pin Set
(Can be connected to the tip of
the L9208, which comes with the
instrument.)
• L4934 Small Alligator Clip Set
(Can be connected to the tip of
the L9208, which comes with the
instrument.)
Basic Specifi cations
Maximum input
current
• Jaw (3280-10F, 3280-20F)
2000 A AC continuous (45 Hz to 66 Hz)
• Flexible loop
(3280-10F+CT6280 or 3280-20F+CT6280)
4200 A AC continuous (50 Hz to 60 Hz)
Maximum input
voltage
600 V AC/DC and 3×10
6
V·Hz or less
(ACV/DCV)
Overload
protection
600 V AC/DC (ACV/DCV/
Ω
/continuity)
Maximum rated
voltage to earth
• Jaw, CT6280
600 V AC (Measurement category III),
300 V AC (Measurement category IV)
(Anticipated transient overvoltage:
6000 V)
• Voltage measurement terminal
600 V AC (Measurement category II),
300 V AC (Measurement category III)
(Anticipated transient overvoltage:
4000 V)