5.2 Function Specifications
32
5.2 Function Specifications
Actual Time
Management
Time management using PC application
Recording
Start/End
Recording is started and ended by a com-
mand. (Immediate start, timed start, or timed
end)
Recording
Method
Interval recording (Measurements are re-
corded at a set recording interval.)
Recorded Data
One data set contains time, temperature,
and humidity (for 1 channel each).
Recording Mode
• Instantaneous value
• MAX/MIN/AVE
• Instantaneous value + MAX/MIN/AVE
Total 3 modes
Set the mode before the start of recording.
Recording End
Condition
Memory full stop or indefinite
Set the condition before the start of record-
ing.
Quantity of
Recorded Data
512 k bytes Flash memory
• Instantaneous value recording mode
30,000 data × 2 CH
• MAX/MIN/AVE recording mode
15,000 data × 2 CH
• Instantaneous value + MAX/MIN/AVE re-
cording mode
12,000 data × 2 CH
Data Acquisition
Method
All logging data, data before, at, and after the
specified time, or the current instantaneous
value (monitored value)
Recording
Interval
1/2/5/10/15/20/30 sec.
1/2/5/10/15/20/30/60 min.
Alarm
Criterion threshold can be set to either Hi or
Lo.
The instantaneous value at every sampling is
judged (effective in any measurement
mode).
Summary of Contents for 2301-20
Page 1: ...INSTRUCTION MANUAL For 2301 20 HUMIDITY MODULE...
Page 2: ......
Page 14: ...Notes on Use 10...
Page 22: ...1 4 Dimension Diagrams 18...
Page 24: ...2 1 Setting the Module ID 20...
Page 34: ...4 2 Insulation of Internal Circuit 30...
Page 40: ...5 4 9764 HUMIDIDTY SENSOR Specifications 36...
Page 42: ...6 2 Service 38...
Page 43: ......
Page 44: ......