3.2 Connecting Input/Output Cables
22
3.2.1 Connecting Sensor Cables to the
Terminal block
Å@
Connect the 9764 HUMIDIDTY SENSOR to the
module by following the procedure below.
1.
Use a flat blade screwdriver to loosen the screws
on the terminal block.
2.
Insert a cable into the terminal block, then tighten
the screws (at a tightening torque of 0.25 N•m).
3.2 Connecting Input/Output Cables
• The TEMP terminal and HUM terminal are not
insulated from each other. Avoid short-circuiting.
• Be sure to connect the cables to matching con-
nectors to prevent damage to the module or
sensor.
Red
Yellow
Green
Blue
Black
White