The MetaNeb® System Instructions for Use
(174432 REV 10)
Troubleshooting and Maintenance
M
AINTENANCE
No preventative maintenance is required for this device.
An extended warranty is available for service related maintenance and
repair. The extended warranty is available for purchase during the first five
years of ownership.
Units returned for maintenance and repair must be handled by
Hill-Rom, and must have a return goods authorization (RGA) number.
For disposal of the controller, return it to Hill-Rom or your Hill-Rom
distributor.
S
HIPPING
AND
P
ACKAGING
When The MetaNeb® System controller is shipped for repair, follow these
shipping and packaging instructions:
1.
Request and get a return goods authorization (RGA) number from
Hill-Rom.
NOTE:
You will get a return kit in the mail.
2.
Clean and disinfect the unit. Make sure it is dry before you pack it. See
“Cleaning” on page 26.
3.
Follow the instructions in the return kit to pack the unit.
4.
Close and seal the box, and apply the applicable labels on the outside
of the box.
5.
During shipment, the unit should be kept dry and maintained at
temperatures of -40° F to 158° F (-40° C to 70° C).
Nebulizer not
aerosolizing
correctly.
Nebulizer connection.
Nebulizer is dirty.
Connect the nebulizer
correctly.
Clean or replace the
nebulizer (see the
Nebulizer package for
cleaning instructions).
Manometer shows
a reading even
though no pressure
is applied.
Adjust the manometer to
zero. Refer to the service
manual.
Problem
Examine
Repair