14
The MetaNeb® System Instructions for Use
(174432 REV 10)
Assemble the Circuit
Connect the Oxygen Hose
1.
Attach, and hand tighten the oxygen hose to the connector on the
controller.
2.
Install the applicable hose adapter on the free end of the hose, as
necessary for your facility.
3.
Attach, and hand tighten the free end of the oxygen hose to an
oxygen source.
ASSEMBLE THE CIRCUIT
T
YPICAL
M
OUTHPIECE
C
ONFIGURATION
1.
Connect the circuit tri-connector/bio-filter to the tri-connector port
on the control unit.
NOTES:
•
Do not apply force or pull the device when it is connected to the
oxygen source.
•
Do not pull the device using the circuit tubing when in use or
during transportation.
2.
Remove the mouthpiece from the package.
3.
Attach the mouthpiece to the handset—insert the mouthpiece at a
45° angle and gently push it in and twist it to the correct orientation.
CAUTION:
Caution
—Use only with the supplied nebulizer.
4.
Remove the nebulizer from the package.
5.
Assemble the nebulizer and add the prescribed medication. See “Add
Medication to the Nebulizer” on page 19.
Summary of Contents for MetaNeb
Page 1: ...The MetaNeb System Instructions for Use Product No PMN4 ...
Page 2: ......
Page 19: ...The MetaNeb System Instructions for Use 174432 REV 10 13 Assemble the Controller System ...
Page 53: ......
Page 54: ......