Liko Ultra LiftPants • 7EN160181-03
Before lifting, remember the following:
• A responsible person within your organization should decide on a case-by-case basis whether one or more
caregivers are needed.
• Ensure that the sling selected meets the patient’s needs with reference to model, size, fabric and design.
• For safety and hygiene reasons, use individual slings.
• Before lifting, plan the lifting operation so that it can be done as safely and smoothly as possible.
• Before lifting, check that the lifting accessory hangs vertically and can move freely.
• Although Liko’s sling bars are equipped with latches, special caution must be exercised: Before the patient
is lifted from the underlying surface, but after the straps have been fully extended, make sure the straps are
properly connected to the sling bar.
• Never lift a patient higher off the underlying surface than is needed to complete the lifting and transfer procedure.
• Never leave a patient unattended during a lifting situation.
• Make sure that the wheels on the wheelchair, bed, gurney, etc., are locked during the lifting/transfer operation.
• Always work ergonomically.
Safety Instructions
Liko’s products undergo continuous development, which is why we reserve the right to make product changes
without prior notice. Contact your Liko representative for advice and information about product upgrades.
Care and Maintenance
Inspect the sling regularly, especially after washing. Inspect the following points with regard to wear and damage:
• Fabric
• Straps
• Seams
• Suspension loops
• Buckles
Do not use damaged lifting accessories.
If anything is unclear, please contact the manufacturer or supplier.
Medical technical class I product
Liko Ultra LiftPants have been tested by an accredited testing institute and comply with all requirements for
MDD Class I products and the harmonized EN ISO 10535 standard.
IMPORTANT!
Lifting and transferring a patient always involves a certain level of risk. Read the instruction guide for both
the patient lift and lifting accessories before use. It is important to completely understand the contents of the
instruction guide. The equipment should only be used by trained personnel. Ensure that the lifting accessories
are suitable for the lift used. Exercise care and caution during use. As a caregiver, you are always responsible
for the patient’s safety. You must be aware of the patient’s ability to make it through the lifting situation.
Washing instructions:
See the sling’s product label.
Protected by Patent.
Liko is quality certified according to ISO 9001 and its equivalence for the medical device industry, ISO 13485.
Liko is also certified according to environmental standard ISO 14001.
Expected Life Time
The product has an expected life time of 1-3 years with normal use. Expected life time varies depending on the
fabric, use frequency, washing procedure, and weight carried.
Periodic Inspection
The product must be inspected at least once every 6 months. More frequent inspections might be required if the
product is used or washed more often than normal. See Liko’s protocol Periodic Inspection for the respective
product.