Service Manual Multirall™
overhead lift
• 3EN250401 Rev. 18
5
www.hillrom.com
2.0 Troubleshooting, Multirall™ 200 overhead lift
The lift doesn’t work.
1. Check that the Emergency Stop button is not
activated.
2.
Check that the battery is charged. If the
battery does not charge check charger
voltage.
3. Check that the hand control is correctly
connected. If connected, check the hoist
switches for operation.
4.
If the lift still does not work satisfactorily,
please contact Hill-Rom.
The lift emits a repeated sound or
light signal.
1.
Charge the battery immediately.
2.
If the lift still does not work satisfactorily,
please contact Hill-Rom.
The lift is stuck in the
high position.
1.
Check that the Emergency Stop button is n
ot
activated.
2. Check that the battery is charged. If the
battery does not charge check charger
voltage.
3. Check that the hand control is correctly
connected. If connected, check the hoist
switches for operation.
4.
Try to lower the lift; the limit switch might
be engaged.
5. Use indicated electrical or mechanical
emergency lowering to lower the patient
onto a firm surface.
6.
If the lift still does not work satisfactorily,
please contact Hill-Rom.
You hear unusual sounds.
Contact Hill-Rom.