2.5 Gas Spring Malfunction
Chapter 2: Troubleshooting Procedures
CarePorter™ Cart Service Manual (man239)
Page 2 - 15
2
2.5
Gas Spring Malfunction
1. Press the rear leg release, and move the CarePorter™ Cart forward. The
gas spring pushes the rear legs back into the locked position.
Yes
No
↓
→
2. Go to “Final Actions” on page 2-5.
3. Inspect the gas spring for proper operation. The gas spring is functioning
properly.
Yes
No
↓
→
Ensure the gas spring is fully extended. Check the gas spring rod
for scratches, nicks, or dents. If necessary, replace the damaged
gas spring (refer to procedure 4.12). If this solves the problem,
go to “Final Actions” on page 2-5. Otherwise, go to step 4.
4. Another part of the CarePorter™ Cart interferes with the gas spring
assembly.
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.
5. Take appropriate actions to eliminate the interference.
This solves the problem.
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.
Summary of Contents for CarePorter P3000
Page 5: ...Revisions Page iv CarePorter Cart Service Manual man239 NOTES...
Page 39: ...Chapter 2 Troubleshooting Procedures Page 2 2 CarePorter Cart Service Manual man239 NOTES...
Page 59: ...Chapter 3 Theory of Operation Page 3 2 CarePorter Cart Service Manual man239 NOTES...
Page 129: ...Warranty Chapter 5 Parts List Page 5 4 CarePorter Cart Service Manual man239 NOTES...
Page 218: ...Column Wiring Diagram Back to Chapter 3...
Page 219: ...Base Wiring Diagram Back to Chapter 3...
Page 220: ...Ventilator Shelf Wiring Diagram Back to Chapter 3...