39
CHANnel VLOw
Command:
CHAN:VLO
a
,
volts
<CR><LF>
Response:
OK<CR><LF>
Query:
CHAN:VLO?
a
<CR><LF>
Response:
volts
<CR><LF>
Description: Sets channel
a
’s VL to
volts
. The possible values for
a
are shown
below.
A
Channel A VL
B
Channel B VL
C
Channel C VL
D
Channel D VL
volts
can be from –5.00 to +4.10 in 0.10 increments. VL must be set at least 0.20
below VH. It is not necessary to include the sign for zero or positive values. The
query response will always include the sign.
Example:
Set Channel A VLO to -2.50V.
CHAN:VLO A, -2.5<CR><LF>
reply:
OK<CR><LF>
Example:
Query Channel A VLO value.
CHAN:VLO? A<CR><LF>
reply:
- 2.5<CR><LF>
CLear Status
Command:
*CLS<CR><LF>
Response:
OK<CR><LF>
Query:
None
Response:
None
Description: This command clears the input serial buffer.
Example:
*CLS<CR><LF>
reply:
OK<CR><LF>