The option in the parameters calibration are as follows:
16.2
16.2
16.2
16.2.1
.1
.1
.1 calibration
calibration
calibration
calibration Index
Index
Index
Index difference
difference
difference
difference
Process as follows: after the index difference correction, the first
pop-up window is "Aim at a target F1", focus on the target, then
press "ENT" to confirm, the instrument will measure the vertical
angle automatically, the reading of vertical angle will stay for 1
second, then the pop-up window will be "Aim at the target F2" ,
press "ENT" to confirm, the instrument measure the vertical
angle and calculates the correction value automatically, and will
show the correction , and ask you whether to save it , press
"ENT button" to save and exit (index difference correction
complete), press " ESC
”
key not to save (index difference will
keep the original value).