PAGE 8
cONTrOlS continued...
language and version information
NOTE: Indicated by a speech symbol to
the left of the display.
NOTE: If the language is left blank (non selected, this is the default)
then the device will scroll through all installed languages.
NOTE: Connecting PurgNet™ accessories, may add further menu
items to this list, as required.
Pressing this button will increase the currently displayed set-point by
10 ppm. Holding this button will cause the set-point to increase in
100 ppm steps. The set-point will not increase beyond the maximum
range of the sensor.
Adjust -
Pressing this button will decrease the currently displayed set-point
by 10 ppm. Holding this button will cause the set-point to decrease
in 100 ppm steps. The minimum set-point is 10 ppm, pressing this
button when 10 ppm is displayed will disable the alarm function.
If you have connected a PurgNet™ accessory with relays included
then two further options will appear, these are as follows:
•
Automatic ALL: Activate the alarm when ALL of the relays are
active.
•
Automatic ANY: Activate the alarm when ANY of the relays are
active.