PAGE 5 

gETTINg STarTED continued

4.  Wait for the sensor to warm up, shown on the display.
5.   Set your inert gas regulator for an outlet pressure of 0.5 bar and 

a flow rate of 7-10 litres/minute. When purging at higher flow 

rates, ensure that there is another exhaust outlet available so 

that all of the exhaust gas doesn’t come through the monitor.

Summary of Contents for PurgEye 1500 Site IP68 WELD PURGE MONITOR with PurgeNet

Page 1: ...USER INSTRUCTIONS PurgEye a Family Series of Weld Purge Monitors PurgEye 1500 Site IP68 with PurgeNet WELD PURGE MONITOR...

Page 2: ...t HFT All rights reserved This publication may not be reproduced by any means without the written permission of Huntingdon Fusion Techniques HFT All logos names trademarks are the property and ownersh...

Page 3: ...ting started 4 5 Connections 6 Controls 7 8 Status icons 9 PurgeNet interface 10 Care and maintenance 11 Warning 12 Other languages 13 Customer support and worldwide care 14 Warranty 15 Address and co...

Page 4: ...the PurgEye 1500 Site IP68 See item 3 below Continue to learn the controls 3 Take the free end of the green tubing and attach it to the white connector on the inside of the instrument The other end of...

Page 5: ...on the display 5 Set your inert gas regulator for an outlet pressure of 0 5 bar and a flow rate of 7 10 litres minute When purging at higher flow rates ensure that there is another exhaust outlet ava...

Page 6: ...ith the sample pump running continuously for up to 10 hours on a simple charge and should re charge in under 2 hours Quick Fit sample gas tube fittings The tube connecting the purge gas being sampled...

Page 7: ...ls NOTE If the alarm becomes inactive due to the current reading dropping below the alarm set point then the mute feature is automatically disabled allowing the sounder to activate next time the alarm...

Page 8: ...cause the set point to increase in 100 ppm steps The set point will not increase beyond the maximum range of the sensor Adjust Pressing this button will decrease the currently displayed set point by...

Page 9: ...plug icon will be displayed when the battery is fully charged Pump status When the sample pump is enabled you will see an arrow icon scrolling to the right if the PurgEye has detected that there is i...

Page 10: ...ies are connected to the PurgeNet interface connection they will be automatically detected by your PurgEye Site and extra menu options may appear as required Language Support Current support The PurgE...

Page 11: ...oved before commencing a measurement Please Note Permanent damage to the sensor can occur due to the following circumstances Physical shock Water vapour condensation etc use water filter Sensor is in...

Page 12: ...ure of the purge gas at 0 5 bar 7 5 psi or lower at all times Calibration HFT has a number of calibration facilities internationally Contact your nearest HFT Distributor who can tell you about your cl...

Page 13: ...n sind momentan im Gange Kontaktieren Sie vorl ufig bitte Ihren rtlichen Verteiler Les autres langues sont actuellement dans le progr s Pour le moment s il vous pla t contacter votre Distributeur loca...

Page 14: ...us at Internet www huntingdonfusion com Email hft huntingdonfusion com Tel 44 0 1554 836 836 Worldwide Offices Partners and Distributors are listed on our website or contact us and we will direct you...

Page 15: ...on Fusion Techniques HFT will not be held responsible for any incorrect use of the product For further warranty information please refer to our terms and conditions All warranties shall not apply to a...

Page 16: ...554 836 837 www huntingdonfusion com Email hft huntingdonfusion com E OE Copyright HFT All rights reserved This publication may not be reproduced by any means without the written permission of Hunting...

Reviews: