Enter setting interface of backlight time. The present interface displays
the backlight time is 10s (100ms as a unit). Press the button "
" and "
" to set the
backlight time among 10~600(i.e. 1s to 60s). After completing setting, press button "
" to
confirm and return to setting interface of address.
7.11 Password Setting Interface
Press buttons "
" and "
" at the same time, it will enter total
parameters setting mode. Enter password setting interface, when setting password as 1111,
press "
" to enter setting interface of energy ratio.
7.12 Energy Ratio Setting Interface
Enter setting interface of energy ratio switch setting interface. Press
button "
" to set the energy ratio, among which 0 signifies “ON” while 1 signifies “OFF”.
After completing setting, press button "
" to confirm and enter setting interface of decimal
places.
7.13 Decimal Place Setting Interface
Enter setting interface of decimal place setting interface. The
present interface displays the 2 decimal places. Press button "
" to set the decimal places
from 1~3. When setting decimal place as 0, 4 or values over 4, there will be an exclamation
mark at the left corner, which means it is unsettable. After completing setting, press button
"
" to confirm and return to password setting interface.
Chapter 8 Basically Measured Parameters Zone (Modbus Protocol)
Address
Address
Parameter Name
Default Value
Value Range
DEC
HEX
4096
1000
phase A voltage
Magnifying 100 times
0
~
65535
4097
1001
phase B voltage
Magnifying 100 times
0
~
65535
4098
1002
phase C voltage
Magnifying 100 times
0
~
65535