19/34
5.3
Enquiry the programme versions
1. Switch off the mains switch.
2. Simultaneously press the pulse key and the arrow key in the speed area.
Switch on the mains switch and release the keys again.
3. Press the key in the speed area so often until the programme version of the
control panel (e.g. b1.12) is displayed.
RPM x 100
START
STO P
t
Speed area
Time area
4. Press the key in the speed area again.
The programme version of the power board (e.g. F1.03) is displayed.
RPM x 100
START
STO P
t
Speed area
Time area
5. To exit the machine version display press the key
STOP
or perform a MAINS-
RESET.
5.4 Brake
setting
With this centrifuge, the braking effect can be set to normal or low.
This can be set before a run, as follows:
1. Switch off the mains switch.
2. Simultaneously press the pulse key and the arrow key in the speed area.
Switch on the mains switch and release the keys again.
3. If necessary press the key so often until the machine version appears in the
speed area and the brake setting appears in the time area.
RPM x 100
START
STO P
t
Speed area Time area
RPM x 100
START
STO P
t
Speed area Time area
Normal braking effect “1”
Low braking effect “0”
4. Set the required value, “1” or “0”, in the time area with the and arrow keys.
For brake times see operating instructions chapter " Anhang/Appendix, Rotoren und
Zubehör/Rotors and accessories ".
5. Confirm the set value with the
STOP
key.