34 / 174
Commissioning
Connection of additional devices
(Switchboard option)
1
2
Fig. 27: Connections for additional
devices
1 Connections for additional devices
2 Access port
NOTICE
An excessively high interior temperature can result in
damage to the incubator.
External devices which are operated inside the incu-
bator must not exceed the permitted power, since the
uncontrolled temperature increase can result in
damage to the incubator.
−
If there is additional heat input in the cooled incu-
bator, observe the maximum possible heat com-
pensation.
−
Monitor the temperature process flow of the incu-
bator.
−
Connect a signal transducer to the potential-free
alarm contact in order to also be informed of fault
conditions outside of the viewing range of the
incubator.
Additional devices which are placed in the incubator, for example,
can be connected here. The outputs can all be switched on and off
together via the control contact
Chapter 6.13.6 ‘Control contact
(option)’ on page 101 of the incubator. The cables of the additional
devices can be guided through the access port
‘Additional device access port’ on page 19 into the interior of the
incubator.
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...