Relative centrifugal force (RCF) 

The relative centrifugal force (RCF) is given as a multiple of the acceleration of gravity (g). It is a unit-free value and 
serves to compare the separation and sedimentation performance. 

These values are calculated using the formula below: 

1,118

r

2

1000

RPM

RCF

 

 

1000

1,118

r

RCF

RPM

 

 RCF  = relative centrifugal force 
 RPM  = rotational speed (revolutions per minute) 
 

r  = centrifugal radius in mm =  distance from the centre of the turning axis to the bottom of the centrifuge 

 

   

container. For more on the centrifugal radius see the chapter 

 

   

“Anhang/Appendix,Rotoren und Zubehör/Rotors and accessories”. 

 

The relative centrifugal force (RCF) stands in relation to the revolutions per minute and the centrifugal 
radius. 

 
 

22 

Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm

3

 

When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not 
exceed 1.2 kg/dm

3

The speed must be reduced for materials or mixtures of materials with a higher density. 

The permissible speed can be calculated using the following formula: 

[RPM]

 

speed

  

maximum

 x 

[kg/dm³]

density 

 

Greater

1.2

 

=

 

)

(n

 

speed

 

Reduced

  

red

 

e.g.: maximum speed RPM 4000, density 1.6 kg/dm

3

  

RPM

 

3464

 

=

 

RPM

 

4000

 x 

kg/dm³

 

1.6

kg/dm³

 

1.2

 

=

 

n

  

red

 

 
If in doubt you should obtain clarification from the manufacturer. 
 
 

23 Emergency 

unlocking 

In the event of a power failure, the lid cannot be unlocked with the motor. Emergency unlocking must be done by 
hand. 

 

To unlock in an emergency, disconnect the centrifuge from the mains. 
Open the lid only when the rotor is at a standstill. 

 

 

CAUTION! Damage to the lock during emergency unlocking by turning the hexagon Allen key in clockwise 
direction (to the right). 

The Allen key may only be turned counter clockwise (to the left); see figure. 

 
 

A

 



Switch off the mains switch (switch setting "0"). 



Look through the window in the lid to make sure that the rotor is at a 
standstill. 



Insert the Allen key horizontally in the bore (A) and turn carefully counter 
clockwise (to the left) until the lid opens. 



Pull the Allen key back out of the bore. 

 

 
 

Summary of Contents for 4050674010009QG

Page 1: ...Rev 02 11 2022 Andreas Hettich GmbH Co KG AB1203EN_SA MIKRO 185 EN Operating Manual 3...

Page 2: ...service hettichlab com www hettichlab com Single Registration Number DE MF 000010680 2015 by Andreas Hettich GmbH Co KG All rights reserved No part of this publication may be reproduced without the p...

Page 3: ...5 16 Entering centrifugation parameters 16 16 1 Direct input of the centrifugation parameters 16 16 1 1 Speed RPM 16 16 1 2 Relative centrifugal force RCF and centrifugal radius RAD 16 16 1 3 Runtime...

Page 4: ...re 32 24 1 2 Surface disinfection 32 24 1 3 Removal of radioactive contaminants 32 24 2 Rotors and Attachments 33 24 2 1 Cleaning and care 33 24 2 2 Disinfection 33 24 2 3 Removal of radioactive conta...

Page 5: ...ocument This symbol refers to important circumstances Symbol on the device and in this document Symbol for the separate collection of electric and electronic devices according to the guideline 2002 96...

Page 6: ...tion in infection diagnostics The centrifuge is only meant for this purpose Another use or one which goes beyond this is considered to be non intended The company Andreas Hettich GmbH Co KG is not lia...

Page 7: ...2450 Nm Obligatory inspection DGUV Regel 100 500 no Ambient conditions EN IEC 61010 1 Set up site Indoors only Altitude Up to 2000 m above sea level Ambient temperature 2 C to 40 C Humidity Maximum re...

Page 8: ...king in a safe and professional manner These operating instructions should be read in conjunction with any other instructions concerning accident prevention and environmental protection based on the n...

Page 9: ...e biosafety systems see chapter Anhang Appendix Rotoren und Zubeh r Rotors and accessories If in doubt you should obtain relevant information from the manufacturer The centrifuge must not be operated...

Page 10: ...and the accessories may only be stored in closed and dry rooms When the device and accessories are stored the following ambient conditions must be complied with Ambient temperature 20 C to 60 C Relat...

Page 11: ...ough there is no fault on the unit Remove the transport securing device from the housing floor see the Removing the transport securing device chapter Position the centrifuge in a stable and level mann...

Page 12: ...A B Installation Clean the motor shaft A and the bore of the rotor and then apply a thin coat of grease to the motor shaft Dirt particles between the motor shaft and rotor prevent the rotor from havi...

Page 13: ...d Centrifugal force When loading the rotor no liquid may enter the rotor or the centrifuging chamber In order to maintain the weight differences within the centrifuge container as marginal as possible...

Page 14: ...value changes with increasing speed SELECT Selection key for activating the individual parameters Every time the key is pressed the next parameter is activated Call up the MACHINE MENU keep the key pr...

Page 15: ...lue can be set which results in a speed between 200 rpm and the maximum rotor speed Adjustable up to 10000 in steps of 1 and above 10000 in steps of 10 It is only possible to input the relative centri...

Page 16: ...the desired value with the keys t min s RPM RCF 16 1 2 Relative centrifugal force RCF and centrifugal radius RAD Example RCF Press the key to activate the RCF display RCF as needed t min s RPM RCF RP...

Page 17: ...parameters must then be input again By pressing the START PULSE button the settings will be saved If several parameters are input the START PULSE key must be pressed after setting the last parameter E...

Page 18: ...et the desired value with the keys t min s RPM RCF t min s RPM RCF SELECT Press the key R Centrifuging radius RCF Relative centrifugal force t min s RPM RCF t Set the desired value with the keys t min...

Page 19: ...saved With the RCF key you can toggle between the RPM display RPM and RCF display RCF at any time If the RCF display RCF is worked with the centrifuging radius must be input The following is displaye...

Page 20: ...y RPM or the RCF display RCF as needed t min s RPM RCF t min s RPM RCF Input the desired centrifugation parameters Set the parameters t min and t sec to zero see the chapter Entering centrifugation pa...

Page 21: ...y and keep it pressed During the centrifugation run the rotor speed or the resulting RCF value and elapsed time are displayed t min s RPM RCF t min s RPM RCF START PULSE Release the key again to end t...

Page 22: ...me the centrifugation run is started and the centrifugation run must be restarted t min s RPM RCF The following is displayed t min s RPM RCF the rotor must be immediately exchanged for a new rotor for...

Page 23: ...t any time by pressing the STOP OPEN key SELECT Press and hold the button for eight seconds t min s RPM RCF SELECT Press the key t min s RPM RCF START PULSE Press the key Device model and its maximum...

Page 24: ...PM RCF SELECT Press the key Frequency converter type Example t min s RPM RCF SELECT Press the key Frequency converter program version Example t min s RPM RCF STOP OPEN Press the key twice to exit the...

Page 25: ...s the key until the following is displayed t min s RPM RCF START PULSE Press the key End beep Acoustic signal after ending the centrifugation run off Signal deactivated on Signal activated t min s RPM...

Page 26: ...rted at any time by pressing the STOP OPEN key SELECT Press and hold the button for eight seconds t min s RPM RCF SELECT Press the key until the following is displayed t min s RPM RCF START PULSE Pres...

Page 27: ...ECT Press the key until the following is displayed t min s RPM RCF START PULSE Press the key t min s RPM RCF t min s RPM RCF SELECT Press the key until one of the following is displayed Lid AutoOpen A...

Page 28: ...M RCF SELECT Press the key until the following is displayed t min s RPM RCF START PULSE Press the key t min s RPM RCF t min s RPM RCF SELECT Press the key until one of the following is displayed Power...

Page 29: ...procedure can be aborted at any time by pressing the STOP OPEN key SELECT Press and hold the button for eight seconds t min s RPM RCF SELECT Press the key until the following is displayed t min s RPM...

Page 30: ...SELECT Press and hold the button for eight seconds t min s RPM RCF SELECT Press the key until the following is displayed t min s RPM RCF START PULSE Press the key Example t min s RPM RCF SELECT Press...

Page 31: ...xtures may not exceed 1 2 kg dm 3 The speed must be reduced for materials or mixtures of materials with a higher density The permissible speed can be calculated using the following formula RPM speed m...

Page 32: ...hrough adhering impurities Ingredients of suitable detergents soap anionic tensides non ionic tensides After using detergents remove the detergent residue by wiping with a damp cloth The surfaces must...

Page 33: ...on Rotors and attachments may no longer be utilised upon indication of wear and tear or corrosion Check the firm seating of the rotor on a weekly basis 24 2 2 Disinfection If infectious material shoul...

Page 34: ...rs must be removed prior to autoclaving Autoclaving accelerates the ageing process of plastics In addition it can cause discolourations in plastics 24 4 Centrifuge containers With leakiness or after t...

Page 35: ...on run The centrifugation run was not finished Open the lid after the rotor is at a standstill Push START PULSE button Repeat the centrifugation run if necessary TACHO ERROR 1 2 Failure of speed impul...

Page 36: ...as Hettich GmbH Co KG in order to provide protection for people the environment and materials it has to be decontaminated and cleaned before being shipped We reserve the right to refuse contaminated d...

Page 37: ...11 x 38 8 x 30 8 x 45 6 x 45 6 x 18 11 x 38 Anzahl p Rotor number p rotor 18 18 18 18 18 18 Drehzahl speed RPM 14000 RZB RCF 7 16654 Radius radius mm 76 76 76 76 76 76 97 sec 16 1 sec 15 Probenerw rmu...

Page 38: ...Drehzahl speed RPM 14000 RZB RCF 7 18845 17749 Radius radius mm 86 86 86 86 86 81 97 sec 15 1 sec 15 Probenerw rmung sample temp rise K 1 22 1 Probenerw rmung bei maximaler Drehzahl und 1 Stunde Lauf...

Reviews: