Hermle Z 130 M Operating Manual Download Page 11

     MAINTENANCE 

 

Z 130 M_V1.15 

© Hermle Labortechnik GmbH 

 

3.

 

MAINTENANCE 

3.1

 

Maintenance and cleaning 

3.1.1

 

 General 

Care: 

Maintenance  of  the  centrifuge  is  confined  to  keeping  the  rotor,  the  rotor  chamber  and  the  rotor 

accessories.  Clean  as  well  the  accessories  especially  the sealings of  the  aerosol-tight  rotors  and 

insert  bolts  of  swing  out  rotors.  Afterwards  lubricate  the  bolts  or  sealings  with  the  recommended 
HERMLE vaseline - order no. 41-5215.  
Lubricants containing molycote and graphite are not allowed.  
Please pay special attention to anodized aluminium parts. Breakage of rotors can be caused even 

by slightest damages. 
In case of rotors, buckets or tube racks getting in touch with corrosive substances the concerned 

spots have to be cleaned carefully. 
Corrosive  substances  are  for  instance:  alkalis,  alkaline  soap  solutions,  alkaline  amines, 

concentrated  acids,  solutions  containing  heavy  metals,  water-free  chlorinated  solvents,  saline 

solutions, e.g. salt water, phenol, halogenated hydrocarbons.  

 

Cleaning – unit, rotors, accessories: 

-

 

Turn the device off and disconnect it from the power supply before you begin any cleaning or 
disinfecting. Do not pour liquids into the housing interior. 

-

 

Do spray disinfectant on the device.

 

-

 

Thorough cleaning not only has its purpose in hygiene but also in avoiding corrosion based on 

pollution. 

-

 

In order to avoid damaging anodized parts such as rotors, reduction plates etc., only pH-neutral 

Detergents with a pH-value of 6-8 may be used for cleaning. Alkaline cleaning agents (pH-value 

> 8) must not be used. 

-

 

After cleaning, please ensure all parts are dried thoroughly, either by hand or in a hot-air cabinet 

(max. Tempe 50°C). 

-

 

It  is  necessary  to  coat  anodized  aluminium  parts  with  anti-corrosion  oil  regularly  in  order  to 

increase their life-spans and reduce corrosion predisposition. 

-

 

Due  to  humidity  or  not  hermetically  sealed  samples,  condensate  may  be  formed.  The 

condensate has to be removed from the rotor chamber with a soft cloth regularly. 

-

 

 

The maintenance procedure has to be repeated every 10 to 15 runs, but at least once at week! 

 
-

 

Connect the unit to the power supply, after the equipment is completely dry. 

-

 

Do not carry out disinfection with UV-, beta- and gamma-raysor other high energy radiation. 

-

 

Metal rotors can be autoclaved. 

-

 

Rotor lid and adapters can also be autoclaved (max. 121°C, 20 min). 

-

 

The tube racks are made of PP and can 

not

 be autoclaved at 134°C. 

 

3.1.2

 

 Cleaning and disinfection of the unit 

1.

 

Open the lid before you turn off the unit. Disconnect it from the power supply. 

Summary of Contents for Z 130 M

Page 1: ... Hermle Labortechnik GmbH Z 130 M _V1 15 Operating Manual for Mini Centrifuge Z 130 M ...

Page 2: ......

Page 3: ... Starting up 5 2 2 Loading rotor 5 Examples of properly balanced rotors 5 Examples for improperly balanced rotors 6 2 3 Power switch 6 3 MAINTENANCE 7 3 1 Maintenance and cleaning 7 3 1 1 General 7 3 1 2 Cleaning and disinfection of the unit 7 3 1 3 Cleaning and disinfection of the rotor 8 3 1 4 Disinfection of PP rotors 8 3 1 5 Glass breakage 9 3 2 Life time of rotors round and rectangular bucket...

Page 4: ......

Page 5: ... different density Hermle centrifuges are intended exclusively for indoor use and for use by qualified personnel Only Hermle original rotors and accessories might be used Any other use or intended use is considered improper From the resulting damage the company Hermle Labortechnik is not liable 1 2 Brief description The unit type Z 130 M is a mini centrifuge which we offer in two voltage variation...

Page 6: ...e designation of the product REF Order number SN Serial number of the product Date of manufacture Manufacturer MAX Drehzahl max allowed speed of the unit KIN EN max kinetic energy with corresponding rotor U I f Allowable voltage max current frequency P Electrical input power before operating read the operating manual Labeling standards and guidelines that are considered ...

Page 7: ... it is not installed correctly Do not operate the centrifuge when dismounted e g without housing Do not run the centrifuge when mechanical or electrical assembly groups have been tampered with unauthorized persons Do not use accessories such as rotors and buckets which are not exclusively approved by HERMLE Labortechnik GmbH except commercially available centrifuge tubes made of glass or plastic D...

Page 8: ...nit according to the standards EN 61010 2 020 Do not place the centrifuge next to a window or a heater where it could be disposed to excessive heat as the performance of the unit is based on an ambient temperature of 23 C 1 5 3 Installation Follow these steps Check whether power supply corresponds with the one named on the manufacturer s rating label which is mounted on the rear panel The line vol...

Page 9: ...ION Never attempt to remove samples until the rotor has come to a complete stop ATTENTION Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical damage Do not operate with extremely corrosive substances which could damage the rotor and unit 2 2 Loading rotor Rotors must be load symetrically and with equal weight see figure 1 The adapter may only be loa...

Page 10: ...130 M_V1 15 6 Examples for improperly balanced rotors 2 3 Power switch The power swich is on the left side at the back wall of the housing see figure 3 Figure 3 Power swich Represents a loaded position Represents an empty position Figure 2 ...

Page 11: ...to the housing interior Do spray disinfectant on the device Thorough cleaning not only has its purpose in hygiene but also in avoiding corrosion based on pollution In order to avoid damaging anodized parts such as rotors reduction plates etc only pH neutral Detergents with a pH value of 6 8 may be used for cleaning Alkaline cleaning agents pH value 8 must not be used After cleaning please ensure a...

Page 12: ...ng of the rotors and accessories set them on a towel Place the angle rotors with bores down to dry them to 5 Dry the rotor cone with a soft dry and lint free cloth and look for damage Do not grease the rotor cone 3 1 4 Disinfection of PP rotors The recommended time for autoclaving 15 20 min at 121 C 1 bar ATTENTION The sterilization time of 20 min must not be exceeded Sterilization again and again...

Page 13: ...ids and caps after five autoclaving cycles Never store aerosol tight rotors or buckets closed 3 1 5 Glass breakage With high g values the rate of glass tube breakage increases Glass splinters have to be removed immediately from rotor buckets adapters and the rotor chamber itself Fine glass splinters will scratch and therefore damage the protective surface coating of a rotor If glass splinters rema...

Page 14: ...aned before the shipment for the protection of persons environment and material Please fill out the below completely form Decontamination certificate at goods return delivery see APPENDIX p III and put the decontamination form to the outside of the box We reserve the right to accept contaminated centrifuges Further on all costs occurred for the cleaning and disinfection of the units will go to the...

Page 15: ...vices in the European Community The disposal of the electrical devices is regulated within the European Community by national regulations based on EU Directive 2002 96 EC pertaining to waste electrical and electronic equipment WEEE In accordance with this any devices delivered after 13 08 2005 on a business to business basis which includes the product may no longer be disposed of in household wast...

Page 16: ... Hermle Labortechnik GmbH Z 130 M_V1 15 12 APPENDIX Table 1 Technical Data I EC conformity declaration II Redemption form Decontamination certificate III ...

Page 17: ...n Ambient conditions EN IEC 61010 1 Environment for indoor use only High Use up to an altitude of 2000 m above MSL Ambient temperature 2 C up to 35 C Max relative humidity Max relative humidity 80 for temperatures up to 31 C Decreasing linearly to 50 relative humidity up to 35 C Overvoltage category IEC 60364 4 443 II Degree of contamination 2 Class of protection I Class of protection DIN EN 60529...

Page 18: ... Product named below fulfills the relevant fundamental requirements of the EC directives and standards listed In the case of unauthorized modifications to the product or an unintended use this declaration becomes invalid Produkttyp Product type Laborzentrifugen mit Zubehör Laboratory centrifuge with accessories Typenbezeichnung Typ designation Z 130 M Einschlägige EG Richtlinien Normen Relevant EC...

Page 19: ...x ____________________ E Mail _____________________________________ Pos Crowd Decontaminated object Serial number Description Comment 1 2 3 4 Are these parts listed above in touch with the following substances Health endangering watery solutions buffers acids alkalis Yes No Potentially infectious agents Yes No Organic reagents and solvent Yes No Radioactive substances α β γ Yes No Health endangeri...

Page 20: ......

Page 21: ...Z 130 M_V1 15 Hermle Labortechnik GmbH 6 Notes ...

Page 22: ... Hermle Labortechnik GmbH Z 130 M_V1 15 ...

Page 23: ...Z 130 M_V1 15 Hermle Labortechnik GmbH ...

Page 24: ...HERMLE Labortechnik GmbH Siemensstrasse 25 78564 Wehingen Tel 0 74 26 96 22 17 Fax 0 74 26 96 22 49 Email vertrieb hermle labortechnik de Internet http www hermle labortechnik de Technische Änderungen vorbehalten HERMLE Labortechnik GmbH 2012 ...

Reviews: