Heraeus PALABOWL Instructions For Use Manual Download Page 2

2

PALABOWL

®

1. Device name

1) Device name: 

PALABOWL

®

 

Bone Cement Mixing System

2) Type: 

CL006-01

3) REF: 5050166

2. Indications

1)  The Bone Cement Mixing System is for mixing of the bone 

cement powder and liquid monomer.

2)  This device can mix up to three units (total 120 g powder) 

of bone cement of any viscosity.

3. Working principles

The Bone Cement Mixing System’s rotating axis paddle allows 
for a thorough cement mix of all angles without unmixed 
powder collecting at the edges of the bowl. While mixing, 
vacuum removes air via a filter from the bowl to reduce 
porosity of the cement and cement fume.

4. Specification

1)  Mixing bowl size (ID×H): 106mm × 72 mm; Vacuum hose 

size (length): 2.4 m.

2)  Net weight: Not more than 0.3 kg.
3)  Air tightness: There is no leak under negative pressure 

75 kPa ± 1k Pa.

4)  Vacuum hose: The vacuum hose shall not shrink under 

negative pressure 75 kPa ± 1 kPa.

5)  Mixing paddle: The Mixing paddle can rotate freely under 

negative pressure 75 kPa± 1 kPa.

6)  Sterilization: Sterilized using ethylene oxide before 

delivery.

7)  Valid period: Two years.
8)  The maximum allowable negative pressure value during 

operating: 73 kPa (550 mmHg).

9)  To reach the negative pressure value 73 kPa (550 mmHg) 

needed time: about 25 s.

5. Components

1)  This device consists of 

Vacuum

 

Mixing bowl, Mixing 

paddle, Spatula, Curette, Vacuum hose

 and 

Rotating 

handle

.

2)  Associated devices: The 

Bone Cement

 and 

Vacuum 

extractor

 are not included in this device.

6. Warning and Precautions

1)  This device shall only be operated by qualified 

professionals.

2)  The bone cement powder and liquid monomer are not 

included in this device but prepared by the hospital.  
The bone cement must be prepared according to the 
instructions for use provided by the bone cement 
manufacturer.

3)  The Vacuum extractor is not included in this device, but 

prepared by the hospital. The negative pressure within the 
range of 5 kPa to 75 kPa shall be adjustable.

4)  This device is sterilized using ethylene oxide before 

delivery. Sterilization validity is two years.

5)  This device is for single patient use only. Do not reuse and 

do not re-sterilize.

6)  Follow current local/national regulations to properly 

dispose the device after use.

7)  During transportation of this device, prevent severe shock, 

vibration, rain/snow, direct sunlight, moist and pressure.

8)  This device should be stored in a non-corrosive, cool, dry, 

and ventilated environment. 

7.   Operating  instructions

1)  Inspect the package prior to each use. Do not use if 

packaging is damaged.

2)  Read the instructions for use, attached documents 

and labels.

3)  Open the package, take out the product and operate the 

device according to following figures:

08611

(6)

(5)

(3)

(4)

(1)

(1)  Mixing bowl
(2)  Mixing paddle
(3) Spatula
(4)  Curette
(5)  Vacuum hose
(6)  Rotating handle

(2)

EN

Summary of Contents for PALABOWL

Page 1: ...Bone Cement Mixing System Instructions for use...

Page 2: ...ng paddle Spatula Curette Vacuum hose and Rotating handle 2 Associated devices The Bone Cement and Vacuum extractor are not included in this device 6 Warning and Precautions 1 This device shall only b...

Page 3: ...the vacuum hose 8 To scrape the bone cement on the mixing paddle into mixing bowl by using spatula or curette 1 Place the bowl on a flat surface and open the lid 3 Pour the single compo nents into th...

Page 4: ...pala mezcladora esp tula cureta tubo de vac o y brazo giratorio 2 Dispositivos asociados El cemento seo y el extractor de vac o no est n incluidos en este dispositivo 6 Advertencias y precauciones 1...

Page 5: ...el vac o y quite el tubo de vac o 8 Raspe el cemento seo de la pala mezcladora en el recipiente de mezcla con la esp tula o la cureta 1 Coloque el recipiente en una superficie plana y abra la tapa 3 V...

Page 6: ...otola di miscelazione sottovuoto paletta di miscelazione spatola curette tubo del vuoto e manovella di rotazione 2 Dispositivi accessori Il cemento osseo e l estrattore sottovuoto non sono inclusi nel...

Page 7: ...sidui di cemento osseo dalla paletta di miscelazione con la spatola o la curette in modo che si raccolgano all interno della ciotola di miscelazione 1 Posizionare la ciotola su una superficie piana e...

Page 8: ...distribution and use by or on the order of a physician Keep dry Do not resterilize 0123 CE marking of conformity Latex free CL006 01 08E Z3 Rev C USA Agent SUNGO Technical Service Inc 6050 W Eastwood...

Reviews: