5.6. Configuration
Select System Configuration through placing the cursor onto
the selected item “system” and operating the ENTER key.
The system configuration menu is displayed on 2 pages.
5.6.1. Configuration System PAGE 1
The language in which the menus are to be displayed is
selected through the language option.
Value range: German / English / French
language is active once it has been selected and acknowledged by pressing the ENTER key.
The menu will be displayed in the selected language.
accesscode - function is used to safeguard write access to the settings. Value range:
off / on
When acknowledging the on setting, the access code entry screen is activated. See
reference (access code Section 7.1.2).
temp. f. comp. - function (temperature path for temperature compensation) is used to
specify which module is to be supply the temperature for the automatic temperature
compensation.
value range - individual / measuring module 1 / measuring module 2 / measuring module
3. Individual setting means that every module carries out its own temperature
measurement for the respective temperature compensation of the measured value.
comparative pH – setting made for pH measurement is responsible for self-monitoring pH
measurements.
value range - no comparison / 2 comparisons / 3 comparisons. The pH module on plug-in
connection 1 must have a controller identifier since only the controller from module 1 is
used.
alarmrel inv - use alarmrel inv. (inverted) to change the mode of operation of the alarm
relay. If the mode of operation is not inverted (setting OFF), the alarm relay is activated in
an alarm scenario; contacts 1 and 2 are closed. If the mode of operation is inverted (setting
ON), the alarm relay is deactivated in an alarm scenario; contacts 1 and 2a are closed. With
this setting, a network failure on the unit can be detected (intrinsic safety).
config. Menu - opens the configuration menu.
Page 2 – function opens the second page of the system configuration menu.
5.6.2. Configuration System PAGE 2
The idle time function is used to specify the time that may
pass between operating two keys before the measurement
screen is automatically opened again.