HemoCue Hb201 Instructions Download Page 3

Limitations of the method

a) The measurement needs to be started no later than 10 minutes after filling the 

microcuvette.

b) Mixing samples for an extended period can produce increased oxygen pressure and 

viscosity that may give falsely results.

c) If “HHH” is displayed, the result exceeds the measuring range of the system. 

d) Values above 23.5 g/dL (235 g/L, 14.6 mmol/L) must be confirmed using a suitable 

laboratory method.

e) Following substances have not been found to interfere: Acetaminophen (20 mg/dL), 

ascorbic acid (3 mg/dL), conjugated bilirubin (40 mg/dL), unconjugated bilirubin 

(20 mg/dL), creatinine (30 mg/dL), ibuprofen (40 mg/dL), leukocytes (600 x 10

/L), 

lipemia (intralipid 4000 mg/L, triglycerides approximately 1000 mg/dL), salicylic acid 

(50 mg/dL), tetracycline (20 mg/dL), trombocytes (2100 x 10

/L), urea (500 mg/dL), 

uric acid (20 mg/dL). The highest concentration or percentages tested is referred to in 

brackets. Intereference studies have been performed according to NCCLS Document 

EP-7

4

.

f)  pH values between 6.3-9.0 do not interfere with the system.

g) Sulfhemoglobin is not measured with this method.

Expected values

2

Children 

11.0-14.0 g/dL (110-140 g/L, 6.8-8.7 mmol/L)

Women 

12.0-15.0 g/dL (120-150 g/L, 7.4-9.3 mmol/L)

Men 

13.0-17.0 g/dL (130-170 g/L, 8.1-10.5 mmol/L)

Children, 2 years to teenage, gradually increase to adult values.

Due to a wide range of conditions (dietary, geographical, etc) which affect normal val-

ues, it is recommended that each laboratory establish its own normal range.

Specific performance characteristics

Within-run and Total precision

Within-run precision was determined according to the NCCLS Document EP5-A

5

The results given below in ”Within-run precision” and ”Total precision” come from 

1

 

batch of HemoCue Hb 201 Microcuvettes and 5 HemoCue Hb 201

+

 Analyzers. No 

recalibration was performed during the analyzing period. Commercially available con-

trols at 2 different levels were used. The hemoglobin concentration was measured in 

duplicate twice a day, morning and afternoon, during 20 consecutive days.

Within-run Precision

Total Precision

Control 

level

N

x– 

g/dL

SD 

g/dL

CV 

%

SD 

g/dL

CV 

%

1

400

7.70

0.057

0.74

0.100

1.30

2

400

15.33

0.078

0.51

0.109

0.71

 

Accuracy

The results of the comparison studies between the HemoCue Hb 201 system and the 

International Council for Standardization in Haematology method (ICSH)

3

 are sum-

marized in the table below. The study was performed on four HemoCue Hb 201

+

  

Analyzer which had not been recalibrated during the study period. 

Study

N

Min 

g/dL

Max 

g/dL

Regression line

Correlation 

coefficient (r)

1

498

4.1

20.5

Y = 1.009X – 0.008

0.998

2

103

9.0

17.7

Y = 0.971X + 0.584

0.934

1 = ICSH (Cyanmethemoglobin method), venous EDTA blood, multicenter study

2 = ICSH (Cyanmethemoglobin method), capillary blood

Summary of Contents for Hb201

Page 1: ...ance and hemoglobin level is calculated The absorbance is directly pro portional to the hemoglobin concentration Principle of the procedure The system consists of an analyzer together with microcuvettes The microcuvette serves both as a pipette and as a measuring cuvette and is for single use only A blood sample of approximately 10 µL is drawn into the cavity by capillary action The analyzer measu...

Page 2: ...ly Always avoid touching the optical eye Do not refill the cavity of the microcuvette Note The microcuvette should be filled whitin 3 minutes after the microcuvette has been taken out of its package When collection tubes or control material are used dispense an aliquot of the well mixed specimen onto a hydrophobic surface Note The microcuvette should be filled whitin 3 minutes after the microcuvet...

Page 3: ... years to teenage gradually increase to adult values Due to a wide range of conditions dietary geographical etc which affect normal val ues it is recommended that each laboratory establish its own normal range Specific performance characteristics Within run and Total precision Within run precision was determined according to the NCCLS Document EP5 A5 The results given below in Within run precision...

Page 4: ...ate Must not exceed the Use by date Use by Year Month Day PATENTED This product is covered by one or more patents Temperature limitation Catalog number Consult instructions for use Bibliography 1 HemoCue Hb 201 Operating Manuals 2 Dacie and Lewis Practical Haemotology Ninth Edition 2001 3 Reference and Selected Procedures for the Quantitative Determination of Hemoglobin in Blood Approved Standard ...

Reviews: