HemoCue Hb201 Instructions Download Page 2

Storage for individually packaged cuvettes

The microcuvettes are to be stored at room temperature (15–30 °C, 59–86 °F) and 

in a dry place. 

Specimen collection and preparation

Capillary, venous or arterial blood may be used. Appropriate anticoagulants (e.g. 

EDTA or heparin) may be used, preferably in solid form to avoid dilutional effects. 

Mix all specimen tubes thoroughly on a mechanical mixer for at least 2 minutes or 

invert the tube 8-10 times by hand. Hemoglobin remains unchanged for days, pro-

vided that the blood does not become infected. If the specimen has been stored in 

a refrigerator, it will be viscid and should reach room temperature before mixing

2

.

Directions for use

Materials required

• 

HemoCue Hb 201 Analyzer

• 

HemoCue Hb 201 Microcuvettes

• 

Lancet (for capillary samples)

• 

Pipette or other transfer device 

(for venous, arterial or control material samples)

• 

Lint-free wipe (non-fraying)

• 

Hydrophobic surface 

(for venous, arterial or control material samples)

Procedure

The operating temperature of the HemoCue Hb 201 system is 15-30 °C (59-86 °F).

Please read the Operating Manual for proper use of the system

1

. See relevant manu-

al for information on repeat capillary sampling. For further information please con-

tact HemoCue.

1. Optical eye

2. Filling end

Remove the microcuvette from the package. Close the vial. Hold the 

microcuvette opposite the filling end and bring into contact with the 

specimen, allowing the cavity to fill completely. Always avoid touching 

the optical eye. Do not refill the cavity of the microcuvette. Note: The 

microcuvette should be filled whitin 3 minutes after the microcuvette 

has been taken out of its package.

When collection tubes or control material are used, dispense an aliquot 

of the well-mixed specimen onto a hydrophobic surface. Note: The  

microcuvette should be filled whitin 3 minutes after the microcuvette 

has been taken out of its package.
Wipe off the outside of the microcuvette with a clean lint-free wipe, 

being careful not to touch the open end. If air bubbles are seen in the 

optical eye of the microcuvette, discard the microcuvette and take a new 

sample. Small air bubbles around the edge do not influence the result.

Place the microcuvette into the cuvette holder and start measurement 

as soon as possible but no later than 10 minutes after filling the micro-

cuvette by gently pushing the cuvette holder to its measuring position. 

Close the cuvette holder. The measuring time is 15-60 seconds for Hb 

values below 20 g/dL. Pull the cuvette holder out to its loading posi-

tion and discard the used microcuvette.

Quality control

The HemoCue Hb 201 Analyzer has an internal electronic selftest. Every time the 

analyzer is turned on, it will automatically verify the measurement performance. 

This test is performed at regular intervals if the analyzer remains switched on.  

Follow local guidelines regarding quality control procedures. If a quality control 

test is required by local or other regulations and therefore should be performed, 

only use controls recommended by HemoCue. For further information regarding 

controls contact HemoCue AB.

Summary of Contents for Hb201

Page 1: ...ance and hemoglobin level is calculated The absorbance is directly pro portional to the hemoglobin concentration Principle of the procedure The system consists of an analyzer together with microcuvettes The microcuvette serves both as a pipette and as a measuring cuvette and is for single use only A blood sample of approximately 10 µL is drawn into the cavity by capillary action The analyzer measu...

Page 2: ...ly Always avoid touching the optical eye Do not refill the cavity of the microcuvette Note The microcuvette should be filled whitin 3 minutes after the microcuvette has been taken out of its package When collection tubes or control material are used dispense an aliquot of the well mixed specimen onto a hydrophobic surface Note The microcuvette should be filled whitin 3 minutes after the microcuvet...

Page 3: ... years to teenage gradually increase to adult values Due to a wide range of conditions dietary geographical etc which affect normal val ues it is recommended that each laboratory establish its own normal range Specific performance characteristics Within run and Total precision Within run precision was determined according to the NCCLS Document EP5 A5 The results given below in Within run precision...

Page 4: ...ate Must not exceed the Use by date Use by Year Month Day PATENTED This product is covered by one or more patents Temperature limitation Catalog number Consult instructions for use Bibliography 1 HemoCue Hb 201 Operating Manuals 2 Dacie and Lewis Practical Haemotology Ninth Edition 2001 3 Reference and Selected Procedures for the Quantitative Determination of Hemoglobin in Blood Approved Standard ...

Reviews: