Helmer Scientific Cart and Pedestal
User Guide
360382/B
8
4. Using two people to lift, carefully place the unit on the pedestal aligning the four threaded holes located where the
leveling feet were with the four open holes on the top of the pedestal. The refrigerator or freezer should be flush with the
front panel and back panel of the pedestal (Figure 9).
Figure 9
5. Using the side panel openings, place a 1/4” lock washer on a 1/4”-20 hex head screw and hand screw the 1/4”-20 hex
head screws (4) in the threaded holes. Install screws in the back threaded holes first and then the front threaded holes.
Figure 10
6. Using a 7/16” wrench or socket, tighten each of the hex head screws (4) in each corner.
7. Replace the side panels and back panel by aligning the holes in the panel with the holes in the base of the pedestal.
Hand thread the screws through the holes and tighten using a #2 Phillips screwdriver.
8. Replace the front cover on the pedestal by aligning the holes on the side tabs of the cover with the holes in the
pedestal. Hand thread the screws through the holes and tighten using a #2 Phillips screwdriver.
Installing Wall Mount Brackets
CAUTION
Wall mount bracket installation is required with pedestal installation to lock the undercounter refrigerator or freezer to the
wall stud and prevent tipping.
1. Align the undercounter wall mount bracket holes with the threaded holes at the top of the undercounter refrigerator
or freezer.
2. Hand thread 3/8"-16 x 1 hex head screws (2) through bracket and threaded holes. Tighten screws using a 9/16" wrench
or socket (Figure 11).
Figure 11
Gen 3