Helmer Scientific Cart and Pedestal
User Guide
360382/B
4
2
Countertop Cart Installation
2 .1
Required Parts
♦
401642-1
Tall Cart
♦
401714-1
Short Cart
2 .2
Required Tools
♦
10 mm wrench or adjustable wrench
♦
#2 Phillips screwdriver
2 .3
Location Requirements
♦
Countertop units installed on a short cart must have minimum 2" clearance above top of unit, and minimum 2" clearance
on each side when placed under counter.
♦ Cart casters must rest on the floor.
NOTICE
The maximum allowable ambient temperature for an under counter installation of the MLR102 is 28°C.
2 .4 Procedure
CAUTION
•
Power the refrigerator off and disconnect AC power before performing this procedure.
• The chamber temperature may increase above the allowable temperature range for stored inventory. Take precaution
to protect items in the chamber from extended exposure to adverse temperature.
• Multiple countertop refrigerators may not be stacked on top of the cart.
• The countertop refrigerator must be installed on the cart per the installation instructions provided.
• Maximum weight per shelf decreases to 10 lbs (4.54 kg) when cart is installed.
Setup
1. Remove or secure moveable pieces such as shelves, probe bottle and probe(s).
2. Carefully position refrigerator to allow access to the underside of the unit.
3. Using your hand, unscrew and remove the 4 leveling feet from the bottom of the refrigerator. Discard or store leveling
feet for later use (Figure 1).
Figure 1