General Information
6 Installation
6.1
Location Requirements
►
Has a grounded outlet meeting national electric code (NEC) and local electrical requirements.
►
Is clear of direct sunlight, high temperature sources, heating vents, and air conditioning vents.
►
Has a minimum of 3” (76 mm) of space behind the freezer for clearance and feature access.
►
Meets the limits specified for ambient temperature and relative humidity.
6.1.1
Placement
WARNING
To prevent tipping:
►
ensure the casters (if installed) are unlocked and the door is closed before moving
the freezer.
►
do not sit, lean, push or place heavy objects on upper door ledge.
1
If casters are installed, install strain relief as shown below to prevent accidental disconnection.
2
Ensure door is closed and casters (if installed) are unlocked.
3
Move freezer into place. Lock casters if installed.
4
Ensure freezer is level.
Strain relief for power cord (for freezers with casters installed)
NOTE
Helmer recommends the use of leveling feet (unless casters are installed) and wall and
floor brackets (PN 400472-2) for stabilization. Contact Helmer Technical Service for parts
and instruction.
6.2
Temperature Probes
For each probe bottle, use:
►
4 oz. (120 mL) of product simulation solution (1:1 ratio of water to propylene glycol or equivalent low-
temperature fluid).