5. Maintenance and Sterilization 

The Helios Rotary Pen consists of four main components.  The cartridge, which is 
purchased separately, pre-sterilized and individually packaged, the handpiece, the 
motor (contained in the upper portion of the handpiece), and the connecting wire.  All of 
the components, with the exception of the cartridge, which are intended for single-use, 

MUST

 be cleaned and disinfected both before and after each individual use.   

The handpiece and the connecting wire must be wiped down with a soft disinfecting 
cloth such as Sani-Cloth, or DEFEND Plus Disinfecting wipes before and after each 
use.  All components of the machine must be covered with a protective covering or 
sleeve during use to reduce the risk of infection. 

The handpiece of the Helios Rotary Pen is made up of two parts.  By holding both the 
upper and lower portion of the machine, and twisting counter-clockwise, the handpiece 
may be unscrewed, revealing the two portions of the handpiece.  The lower portion may 
be 

DRY HEAT autoclaved

 for sterilization.  You cannot sterilize any part of the Helios 

Rotary Pen by 

STEAM AUTOCLAVE!

  Please be advised that the aluminum body has 

been anodized.  Over time, if you continue to autoclave this portion of your machine, it 

MAY

 cause the color to begin to fade.  This is completely normal and is 

NOT

 covered 

under the warranty.   