HelioGD MAP50 VER. 1.0
18
For the latest documentation please visit
or email [email protected]
S
PAN
C
ALIBRATION
T
RANSMITTING TO DETECTOR
Calibration Menu
Choose Detector:
1
Gas:
Oxygen
Set Zero:
0.0
>>
Set Span:
1.00
Transmit:
Complete
Reading:
20.9%Vol
M=Exit > = Scroll Up/Down Adjusts
Press to select option
Press or
to choose the relay
Press
to go back
Adjust the Span to the desired value
After transmitting the value, the
detector will save it and the reading
will be updated accordingly.
Calibration Menu
Choose Detector:
1
Gas:
Oxygen
Set Zero:
0.0
Set Span:
1.00
Transmit:
Transmitting…
Reading:
20.9%Vol
M=Exit > = Scroll Up/Down Adjusts
Press to select option
Press or
to choose the relay
Press
to go back
The Zero & Span values are sent to
the detector which saves them and an
updated reading in achieved.