Light exposure hazard

Because prolonged intense light exposure can damage the retina, the use of the device for ocular 
examination should not be unnecessarily prolonged, and the brightness setting should not exceed 
what is needed to provide clear visualization of the target structures.
The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure 
time. If the value of radiance were reduced in half, twice the time would be needed to reach the 
maximum exposure limit.
While no acute optical radiation hazards have been identified for the funduscope, it is recommended 
that the intensity of light directed into the patient’s eye be limited to the minimum level which is 
necessary for diagnosis. Infants, aphakes and persons with diseased eyes will be at greater risk. 
The risk may also be increased if the person being examined has had any exposure with the same 
instrument or any other ophthalmic instrument using a visible light source during the previous 
24 hours. This will apply particularly if the eye has been exposed to retinal photography.

Disposal

The product must be recycled as separated electrical and electronic devices. Please observe the 
relevant state-specific disposal regulations.

The appendix contains following tables
- Electromagnetic disturbances – Requirements and tests 
- Technical specification 
- Explanation of the used symbols

Summary of Contents for iC 2

Page 1: ...HEINE iC2 Funduscope V 200 00 069 HEINE Optotechnik GmbH Co KG Kientalstr 7 82211 Herrsching Germany Tel 49 0 8152 38 0 Fax 49 0 8152 38 202 E Mail info heine com www heine com MED 113539 2017 12 15 ...

Page 2: ...on 2 Double stage release button for switch on the device and for picture recording 3 Micro USB port 4 Charger indicator 5 Data matrix code and serial number of the device 6 Mounting case for the smartphone 7 Protective cover Setting up Installation of the app Install the iC2 app on your smartphone The app is available in the AppStore To avoid any possible malfunctions please use the smartphone wi...

Page 3: ...wing procedure Closing the iC2 app causes the device to switch off automatically If there is no Bluetooth connection between the iC2 and the app for five minutes and the double stage release button isn t pushed the device will switch off automatically Reset the device If the device does not react it can be reset by pushing the release button for 30 seconds This procedure has no influence on the da...

Page 4: ... more information please visit www heine com The expected life cycle when the device is normal used and the warning and safety information as well as the maintenance instructions are observed is up to 5 years Beyond this period the product may continue to be used if it is in a safe and good condition General warnings Use only CE marked USB power supplies 5 V from reputable manufacturers that confo...

Page 5: ... funduscope it is recommended that the intensity of light directed into the patient s eye be limited to the minimum level which is necessary for diagnosis Infants aphakes and persons with diseased eyes will be at greater risk The risk may also be increased if the person being examined has had any exposure with the same instrument or any other ophthalmic instrument using a visible light source duri...

Reviews: