GAMMA GP
Press the button to close the valve completely (1).
Pump the cuff by pressing the bulb and release the
air from it by opening the valve. (1+2)
Please note: It is not possible to pump up the cuff
with the button in position (2).
Hygienic Reprocessing
The instruction is available:
- at www.heine.com
- in a paper version which you can request from the address listed
Maintenance
The device is maintenance free.
Service
Note technical inspection:
A technical inspection should be carried out at the latest every two years by
the manufacturer or an authorized service technician.
The referenced measuring device must be traceable to a national/international standard.
A rectangular label is attached to the back of the instrument which displays a year-mark. This indicates
the year in which the instrument should be reinspected.
General Notes
The warranty for the entire product is invalidated if non-genuine HEINE products or non-original parts are
used and if repairs or modifications are made to the device by persons not authorized by HEINE.
For more information, please visit www.heine.com.
The expected life cycle, when the device is normal used and the warning and safety information as well
as the maintenance instructions are observed, is up to 7 years for G5 and GP and up to 12 years for G7
and XXL LF.
Beyond this period, the product may continue to be used if it is in a safe and good condition.
All components of this sphygmomanometer are latex-free, the inflation bulb is also phthalate-free.
Please refer to the enclosed assembly instructions.
General Warnings
Check the correct operation of the device before use! Do not use the device if there are visible signs of
damage.
Do not modify the device.
Use only original HEINE parts, spare parts, accessories and power sources.
Repairs shall only be carried out by qualified persons.
Protect the instrument from impact or shocks (do not drop) and protect from moisture and dirt.
Avoid damaging the cuff, tube and bulb with pointed objects such as needles or scissors.
Do not expose the sphygmomanometer to direct sunlight for long periods of time.
This product is not allowed to enter or be used in areas with strong magnetic fields e.g. MRI scanners.
Disposal
Disinfect the product before disposal. Please observe the relevant state-specific disposal regulations.