98
2 Installation and Specifications
2.3 Sof
tw
a
re
set
u
p
Stage squareness calibration
The Squareness screen contains data and choice fields for calibrating
the squareness of the measuring system. The calibration of stage
squareness requires the use of a certified square artifact.
To calibrate squareness:
U
Align the squareness calibration artifact with the reference axis for
the squareness calibration. This axis will be the Master axis.
U
Press MENU/SETUP to display the setup menu and then highlight
the Squareness menu item.
U
Highlight the Master Axis choice field and press a soft key to choose
the reference (Master) axis for squareness calibration.
U
Highlight the Angle data field and press the TEACH soft key to begin
the squareness calibration.
U
Follow instructions displayed on the LCD screen to complete the
calibration.
U
Press the FINISH key to save parameters and return to the setup
menu.
Squareness calibration is unnecessary if optional NLEC
error correction will be used. Error correction is discussed
later in this chapter.
Choose a Master Axis for the calibration Press the TEACH soft key to begin the